System Initialization

System initialization prepares the system for data acquisition. On initial startup, the System
Initialization should consist of:
• Fluidics Prep

• Calibration

• Performance Verification
Calibration normalizes the settings for the system and ensures optimal and consistent
microsphere classification. Verification uses system controls to ensure that the analyzer is
functioning properly with current calibration settings.
1. On the Admin page, select the System Setup tab.

2. In the Maintenance Options section, select the System Initialization procedure that

contains Fluidics Prep, Calibration, and Performance Verification.

3. Click Save.

4. On the Home page, click System Initialization under Daily Activities. The Auto Maint

tab opens.

5. Import the Calibration Kit lot information from the CD provided with the kit or select the

appropriate kit from the drop down menu if the kit information has been preloaded.

NOTE: See the

Adding or Importing CAL and VER Kit Information

section on

how to import the kit.

6. Import the Performance Verification Kit lot information from the CD provided with the kit

or select the appropriate kit from the drop down menu if the kit information has been
preloaded.

NOTE: See the

Adding or Importing CAL and VER Kit Information

section on

how to import the kit.

7. Vortex the xMAP calibrator, verification, and fluidics containers at a medium speed for

approximately 10 seconds to ensure homogeneity. Do not dilute xMAP calibrator,
verification, or fluidics agents.

8. Click Eject on the status bar.

NOTE: To ensure that you get the necessary bead count, invert the

calibrator and verifier vials perpendicular to the strip well as you add
drops to the wells. This ensures that the maximum fluid drop size is
dispensed into the wells.

9. Luminex recommends adding 6 drops of each reagent into the designated well.

10. Verify that reservoir RB1 is 3/4 filled with 70% isopropanol or ethanol.

NOTE: The rinse reservoir (RD1) should be empty.

11. Click Retract.

12. Click Run.