Daily start-up activities, Defining the system initialization routine, Running the system initialization routine – Luminex xPONENT for MAGPIX IVD User Manual

Page 15: Chapter 2 daily start-up activities, Chapter 2: daily start-up activities

Advertising
background image

Defining the System Initialization Routine

NOTE: Luminex recommends daily performance verification and weekly

calibration of the MAGPIX system. You can set up the system
initialization routine to include calibration and verification under the
Admin page, System Setup tab, Maintenance Options section.

1. Open the Admin page.

2. Click System Setup.

3. Under Maintenance Options, select Fluidics prep, performance verfication from the

drop-down list.

FIGURE 3.

Setting System Initialization Routine

4. Click Save.

Running the System Initialization Routine

Chapter 2: Daily Start-Up Activities