Analysis settings dialog box – Luminex xPONENT 3.1 Rev 2 User Manual

Page 39

Advertising
background image

not enough bead events were acquired. If you select bead sets that are not present, the
analyzer continues to acquire, trying to reach the number of events per bead for bead sets
that are not in the sample. Therefore, choose only the bead sets present in your sample.

Apply All - Applies the information in the Units and Counts boxes to all analytes.

Total Count - Select Stop after bead count reaches: to stop acquisition when the bead
count reaches a certain number determined by the user. Type the desired value in the box.
The default value is 100.

You can also specify the minimum allowable bead count per well that the xPONENT

®

software analyzes. This excludes data from any beads carried over during acquisition. For
more information about this feature, see the Batch Options Tab section

Selected Analytes List - Selected analytes appear in a list on the right side of the analyte
grid. This list includes the following information

Name - The name of the analyte. Click and type to rename the analyte.

Analysis - To change the type of analysis for an analyte, click this field to open the

Analysis Settings dialog box and select another analysis from the list.

In the Analysis Settings dialog box:

FIGURE 12.

Analysis Settings Dialog Box

xPONENT

®

3.1 Rev. 2