Results Tab

FIGURE 20.

Results Tab

The Results tab has the same Save Image, Progress, and Well Report buttons as the
Current Batch tab. For more information about these buttons, see

Current Batch Tab

. The

following buttons are added to the Results tab:

Formula - Opens the Change Analysis dialog box with a list of analytes used in the batch.

Click an analyte to open the Analysis Settings dialog box from which you can select a
new analysis setting for the analyte. For more information, see

Create a Multi-Batch

.

Approve - Opens the Batch Approval Confirmation dialog box, which contains the data

for analytes selected in the Results tab. Click Yes to approve the batch. The dialog box
confirms the approval.

Validate - Validates an entire selected row or cell in the Results table. Average rows or

cells cannot be selected. If you haven’t selected an item or the item you selected does not
need to be validated, a warning dialog box displays.

Invalidate - Invalidates an entire selected row or cell in the Results table. The selection

will turn red when invalidated. Select the same item and click Validate to remove the
invalidation status.

Analyze - Analyzes data that has been invalidated. This button only appears if you have

clicked Invalidate on a portion of the data, or if you have clicked Validate on an
invalidated portion of the data. This will analyze the data again with the invalidated
standard removed.

Close - Closes the batch and reopens the Saved Batches tab.