Luminex xPONENT 3.1 Rev 2 (IVD) User Manual

Page 58

Advertising
background image

FIGURE 32.

Maintenance Commands and Routines Tab

Routine Name - This list of pre-set routines and commands can be used for system
maintenance. Some of these commands are also available on the Auto Maint tab.
Additionally, you can create routines on this tab, which display in the Routine Name list once
you save the routine. To create a routine, start with None and add the desired commands.
You can also modify a Luminex routine and save it as a new routine. Select from the
following routines:

Alcohol Flush (Luminex)

Daily Instrument Shutdown (Luminex)

Daily Instrument Startup (Luminex)

Fluidics Maintenance (Luminex)

None

Prepare for Storage (Luminex)

Remove Air Bubbles (Luminex)

Remove Clog (Luminex)

Revive After Storage (Luminex)

Sanitize(Luminex)

Weekly Maintenance (Luminex)

Commands - This section contains the following commands:

Wash - Sends distilled water through the fluidics lines in the system. It pulls the fluid from a

well or from the reservoir and runs it completely through the system to the waste
receptacle.

xPONENT

®

3.1 Rev. 2 Software User Manual

48

For In Vitro Diagnostic Use

Advertising