3. Click OK to start the restore. A progress bar displays, followed by a Warning dialog box

stating that xPONENT needs to be restarted.

4. Click OK to restart xPONENT.

Licensing

You can use the functions on the Licensing tab to change your license information if you are
upgrading to a different software package (from Basic to Secure, for example). If you are
using a trial license and it has expired, contact Luminex technical support.

To obtain a new license:

1. Click Export. The Save As dialog box opens.

2. Select the folder in which you want to save the license file, then click Save.

3. Send an email, with the license file attached, to Luminex Technical Support

([email protected]). Please include the analyzer serial number in the body of
the email. Tech Support needs this in addition to the license key in order to generate a
permanent license.

4. Luminex Technical Support will create a new license file and E-mail it to you. Save the

new license file to your PC or other storage device.

To use the new license:

1. Click License. The Change License dialog box opens. Click Browse, navigate to the

new license file, and then click Open.

2. Read the license agreement.

3. Select Accept License Agreement at next start, then click OK.

NOTE: You MUST restart the computer for the license to take effect.

Schedule

Use the Schedule tab to enable or change the recurrence of scheduled maintenance
reminders.

1. Select the desired reminder time in the Alert Time list.

2. Select how often you want the reminder with the Recurrence list.

3. Enable or disable the reminders by checking or clearing the Enabled check box.

Adjusting the Sample Probe

Adjust the sample probe to ensure that it drops far enough into the well to acquire sample
without impacting the plate.

NOTE: Ensure that there is no liquid in the wells or reservoirs.

NOTE: You should adjust the probe height using row 1 unless you are using

the Automated Maintenance Plate. In that case, the probe
automatically moves to the specified well.

1. On the Home page, click Probe & Heater. The Probe & Heater tab opens.

xPONENT

®

3.1 Rev. 2 Software User Manual

74

For In Vitro Diagnostic Use