FIGURE 47.

Cmds and Routines Tab

Routine Name - This list of pre-set routines and commands can be used for system
maintenance. Some of these commands are also available on the Auto Maint tab.
Additionally, you can create routines on this tab, which display in the Routine Name list once
you save the routine. To create a routine, start with None and add the desired commands.
You can also modify a Luminex routine and save it as a new routine. Select from the
following routines:

Alcohol Flush (Luminex)

Daily Instrument Shutdown (Luminex)

Daily Instrument Startup (Luminex)

Fluidicis Maintenance (Luminex)

None

Prepare for Storage (Luminex)

Remove Air Bubbles (Luminex)

Remove Clog (Luminex)

Revive After Storage (Luminex)

Sanitize (Luminex)

Weekly Maintenance (Luminex)

xPONENT for MAGPIX