System monitor – Luminex xPONENT for MAGPIX User Manual

Page 47

Advertising
background image

Copy All - Copies all data

Copy - Copies only selected data

Paste - Pastes previously copied text or data into the box

Delete - Erases text or data from the selection

System Monitor

The System Monitor displays at the bottom of all xPONENT windows. It displays the
physical state of the Luminex system. Values are reported directly from the Luminex system.

FIGURE 15.

System Monitor

1. System Status Button 2. Connection Status

3. Command Display

4. Progress Bar and Buttons

5. Eject XY Button

6. Drive Fluid Level

7. Waste Fluid Level

8. Delta Cal Temperature

9. XY Status

10. Power Off Button

System StatusButton - This button has two functions: When clicked, it opens the system
log. It also displays the current status of the system. If there are no warnings or errors, the
System Status button is green with a check mark. If there is a warning or other important
user notification, the button is yellow with an X.

Connection Status - Displays the status of the analyzer’s connection to the PC (Connected
or Disconnected). To ensure the analyzer connects to the PC, turn on the analyzer before
you start xPONENT.

Command Display - Displays the following:

• The command currently running.

• The system state (i.e. running, idle, etc.).

• Date and time.

A yellow Check Calibration button is visible in the command display if the calibration or
verification command has failed, when any verification is not current, or when the calibration
or verification was performed prior to the calibration expiration setting.

Progress - Displays a bar graph showing the progress of the current command or routine; if
the command or routine is finished, it displays a full progress bar and the command status as
Complete.