Luminex xPONENT for MAGPIX User Manual

Page 65

Advertising
background image

FIGURE 24.

Settings Tab

This tab contains the following:

Name and description boxes - Type the name and description in the appropriate box.

Acquisition Settings: -

Volume - This is the volume the instrument will aspirate into the system for analysis.

XY heater - Select to enable the XY heater. In the box type the desired value in degrees

Celsius. The temperature range is 35 to 60° C.

CAUTION:

Acquiring data before the heater has reached the proper
temperature can compromise test results.

Plate name - The name assigned to the plate during sample probe height adjustment.

Sample wash - This automatically washes each sample within the instrument.

Analysis Settings - Displays the analysis type to be used for the batch.