Perma Pure MD-Series User Manual

Reflux method, Warning

Advertising
background image

Figure 1 - Typical Setup

2. REFLUX METHOD

The reflux setup, shown in Figure 2, returns

dry sample back to the dryer for use as the

purge after it has gone through the analyzer.

Since this method uses all of the dry sample

as the purge gas, only the sample flow

required for analysis passes through the
dryer. This results in high drying efficiency

.

The vacuum on the purge gas should be at

least 15” of Hg, with a higher vacuum

preferable. This vacuum level is necessary to

provide the desired 2:1 purge-to-sample flow

ratio based on the actual volumetric flow.

Figure 2 - Reflux Setup

The split sample method, shown in Figure

3, diverts some of the sample from the main

stream to be used as the purge gas. More

sample passes through the dryer than is

required for the analysis, lowering the

drying efficiency somewhat.

The following equation can be used to

determine the purge flow rate required for

the split sample method. Any units may be

used as long as they are consistent.

NOTE: Pressure units must be in absolute

terms.

V

p

= V

s

(P

s

/2P

v

) - 1

Where:

Vp = Purge flowrate (indicated on flowmeter)

Vs = Sample flowrate (indicated on flowmeter)

Ps = Sample pressure (in absolute units)
Pv = Purge pressure (in absolute units)

Figure 3 - Split Sample Setup

INSTALLATION SPECIFICATIONS

1. STANDARD METHOD

3. SPLIT SAMPLE METHOD

PRINCIPLE OF OPERATION

MD™-Series gas dryers are shell and tube

moisture exchangers that transfer water vapor

between two countercurrent flowing gas

streams. The dryers consist of a Nafion

®

polymer tube surrounded by an outer tube.

Dry purge gas flowing over the exterior surface

of the Nafion tubing continuously extracts

water vapor from the gas stream inside the

tubing. The driving force is the difference in

water concentration on the opposite sides of

the tubing wall. The purge gas then carries

the water vapor away.

The most efficient way to set up MD-Series

dryers is to have sample enter through the

Nafion tube (wet sample inlet) and purge gas

flowing countercurrent to the sample (refer to

Figure 1). Purge gas should be instrument

air or other dry gas. If no dry purge air is

available, one of the following methods may

be used.

When installing MD-Series gas dryers, the

following rules apply:

1. Sample pressure equal to or greater than

purge pressure

2. Sample gas pressure not to exceed 80 psig

3. Temperatures must not exceed 120°C

4. Purge air of -40°C dew point at a flow rate

of two to three times sample flow

5. Sample and purge air must flow counter-

current to each other

6. If sample dew point is above ambient

temperature, inlet of dryer must be heated

(contact factory for details)

Adjusting end fittings without following the

steps on the back page will cause twisting

of the membrane tubing and void the

warranty.

WARNING!

EXHAUST

VACUUM PUMP

NEEDLE

VALVE

VACUUM GAUGE

GAS

ANALYZER

FLOW

METER

DRY

SAMPLE

OUTLET

WET

SAMPLE

INLET

MD DRYER

EXHAUST

VACUUM PUMP

VACUUM GAUGE

GAS

ANALYZER

DRY

SAMPLE

OUTLET

WET

SAMPLE

INLET

PUMP

(OPTIONAL)

(OPTIONAL)

PURGE

AIR

FLOW

METER

NEEDLE

VALVE

SAMPLE

FLOW

METER

NEEDLE

VALVE

PRESSURE

GAUGE

EXHAUST

MD DRYER

NEEDLE

VALVE

GAS

ANALYZER

MD DRYER

FLOW

METER

DRY

SAMPLE

OUTLET

WET

SAMPLE

INLET

INSTRUMENT

AIR

INLET

DRY

SAMPLE

EXHAUST

PURGE

EXHAUST

PRESSURE

REGULATOR

Advertising