PRINCIPLE OF OPERATION

The ACES system incorporates a Perma Pure Nafion

®

membrane dryer,

disposable 0.1 micron pre-filter and sample pump to supply a clean sample to an
electrochemical sensor. The dryer is sized to supply a 20-80% RH sample from
ambient air at a flowrate up to 2 L/min.

After the ACES provides clean sample to the analyzer, it flows back to the
conditioning system where it is expanded by a vacuum pump and used as purge
air for the dryer (refer to flow schematic below).

INSTALLATION

Mount system close as possible to probe and analyzer to reduce response
time and flow restriction. Enclosure is NEMA 4X rated. Since unit is not
heated, dewpoint of sample entering system can be no higher than the
ambient temperature.

1. Turn power on. Pump will start and sample gas will begin to flow.
2. Adjust valve on top of flowmeter to set desired sample flow rate. Optimal

operating range is 0.5 to 1.5 liters per minute. Typically, a lower flow rate of
sample gas will be dried to a lower dew point than a high sample gas flow
rate. Refer to MD dryer performance chart.

OPERATION

Filter Replacement:

To determine if filter needs to be replaced, monitor pressure drop and flow
output of pump. If output begins to drop, it indicates clogged filter element.

1. Take notice to orientation of filter for replacement. Flow direction is top to

bottom.

2. Pull the lock ring on black elbow fitting away from filter to release internal

gripper. A small adjustable cresent wrench is helpful. Adjust it around the
filter tube stub and slide it down to press the lock ring and then pull the
filter out.

4. Replace with new filter (part number: AG-DIF-K70).

Pump:

The pump is warranteed against failure for 10,000 hours of continuous

operation from Hargraves (model BTC™). If pump fails, it will be repaired or

replaced at manufacturers discretion free of charge. The pump is not

serviceable after warranty period and must be replaced.

MAINTENANCE

SPECIFICATIONS

1. Mount system to wall using the supplied mounting feet. the flow meter must be
in the vertical orientation for proper operation.
2. Connect the fitting labeled "Wet Sample Inlet" to the tube from sample point.
3. Connect the fitting labeled "Dry Sample Outlet" to the tube from the analyzer
inlet.
4. Connect the fitting labeled "Dry Sample Return" to the tube from the analyzer
exhaust.
5. The fitting labeled "Sample Exhaust" can be connected to a vent line if the
sample gas is hazardous,corrosive,toxic, etc.This is not required for proper opera-
tion.

Description

Specification

Dimensions

12"W x 7"H x 6"D

Electrical Requirement

110VAC/2A or 220 VAC/1A

Sample Flow

0.1 - 2.0 L/min

Disposable Filter

95% of 0.1 micron

Inlet Humidity Range

0-99%RH

Outlet Humidity

0-80%RH
1/4" O.D. tube,
polypropylene compression fitting

Filter Model

AG-DIF-K70

Dryer Model

MD-110-12F-4

Piping Connection

Sample Flowmeter

Sample Prefilter

Power Switch

Fuse

Power
Connection

Fiberglass
Enclosure

Sample Gas Connections

SAMPLE EXHAUST

F1

SAMPLE INLET

PARTICULATE

FILTER

DRY SAMPLE RETURN
FROM SENSOR

PUMP

DRY SAMPLE OUTLET TO
SENSOR

NAFION MEMBRANE DRYER

FLOW METER

WITH VALVE