Perma Pure ACES Eductor version User Manual

Principle of operation, Operation, Maintenance installation specifications

Advertising
background image

PRINCIPLE OF OPERATION

The ACES system incorporates a Perma Pure Nafion

®

membrane dryer, disposable

0.1 micron pre-filter and a vacuum eductor to supply a clean sample to an electro-
chemical sensor. The dryer is sized to supply a 20-80% RH sample from ambient
air at a flowrate up to 2 L/min.
Refer to the flow sechematic below. By generating a vacuum, the eductor acts as a
pump to draw the sample gas through the system. The sample gas is first pulled
through the filter to remove particulate matter.Then it flows though the inside of the
Nafion membrane tube within the dryer assembly where excess water is removed.
The clean and dry sample gas then exits the system and flows to the gas sensor
where the measurement is taken. The sample gas must now flow back to the
system to be used as a purge gas for the membrane dryer. Upon re-entering the
system it passes through the flowmeter and through a needle valve where it is
expanded to the vacuum level of the eductor. The expanded sample gas is then
suitable use as purge gas and is passed through the annular area of the dryer
assembly and is exhausted to the atmosphere.

1. Mount system to wall using the supplied mounting feet. the flow meter must be
in the vertical orientation fro proper operation.
2. Connect the fitting labeled "Wet Sample Inlet" to the tube from sample point.
3. Connect the fitting labeled "Dry Sample Outlet" to the tube from the analyzer
inlet.
4. Connect the fitting labeled "Dry Sample Return" to the tube from the analyzer
exhaust.
5. Connect the fitting labeled "Compressed Air Inlet" to the compressed air supply

Mount system close as possible to probe and analyzer to reduce response time
and flow restriction. Enclosure is NEMA 4X rated and can be installed outdoors.
However the unit should be protected from direct exposure to the elements.

1. Turn compressed air on. Eductor will draw sample through the system.
2. Adjust regulator to 10-15" Hg of vacuum. Higher vacuum will provide dryer

sample.

3. Adjust valve on top of flowmeter to set desired sample flow rate. Optimal

operating range is 0.5 to 1.5 liters per minute.

OPERATION

Filter Replacement:

To determine if filter needs to be replaced, monitor flow rate of sample gasof
pump. If output begins to drop, it may indicate a clogged filter element.

1. Note orientation of filter. Flow direction is upward.
2. Push lock ring on black elbow fitting twards body of the fitting. This will

retract the internal gripper that holds the filter in the fitting.

3. Pull tube stubs from fitting and repeat for the other end
4. Replace with new filter (part number: AG-DIF-K70).

MAINTENANCE

INSTALLATION

SPECIFICATIONS

Sample
Flowmeter

Sample
Prefilter

Dry Sample
Outlet

Dry Sample
Return

Wet Sample
Inlet

Compressed
Air Inlet

Compressed
Air Exhaust

Fiberglass
Enclosure

Pressure
Regulator

SAMPLE EXHAUST

F1

SAMPLE INLET

PARTICULATE

FILTER

DRY SAMPLE RETURN
FROM SENSOR

PUMP

DRY SAMPLE OUTLET TO
SENSOR

NAFION MEMBRANE DRYER

FLOW METER

WITH VALVE

Description

Specification

Dimentions

12"W x 7"H x 6"D

Air Supply Pressure Minimum of 60 psig
Air Usage

1.5 - 2.5CFM

Sample Flow

0.1 - 2.0 L/min

Disposable Filter

95% of 0.1 micron

Inlet Humidity Range 0-99%RH
Outlet Humidity

0-80%RH
1/4" O.D. tube,
polypropylene compression fitting

Filter Model

AG-DIF-K70

Dryer Model

MD-110-12F-4

Piping Connection

Advertising