Board Configuration

The board operation is controlled through the bank of DIP Switches labeled SW1 located
on the component side of the board in the lower left corner. These switches must be set
for the cooler to operate properly. The switches numbered 1, 2, and 3 of the DIP switch
set the function of the board.

SW 1

SW2

SW3

Single Channel Cooler with communications

OFF

OFF

OFF

Single Channel Cooler 2

nd

board

ON

OFF

OFF

1

st

Frozen

channel

OFF

OFF

ON

2

nd

Frozen channel

ON

OFF

ON

Dual

Channel

Cooler

OFF

ON

OFF

NJ

TC

Display ON

ON

OFF




Display

Single Channel Cooler

Basic Display Description

3 digit plus sign temperature display
Temperature is displayed in tenths of °C

Green LED to display ‘Ready’ status

Green LED to display ‘Water Slip’ status

Red LED to display ‘TC Fail’ status


Detailed Display Description

The digital display shows both temperature and status of the cooler.

The display will display temperature constantly when the impinger
temperature is below the user set ‘Ready’ temperature and no other faults
are present. One fault is the ‘Water Slip’, the sensing of liquid water
downstream of the sensor, though the use of a water slip sensor. The other
fault would be a ‘TC Failure’ indicating an open thermocouple sensor.

The display will blink the temperature when impinger temperature is
above the user set ‘Ready’ temperature and no other faults are present.
The display switches every second between blank and the temperature, in
this condition.

If a ‘Water Slip’ failure is detected the display will switch between
showing the measured temperature and a symbolic “SLP”. The display
switches every second between “WS” and the temperature.

If a ‘TC Failure’ is detected the display will switch between showing
blank and the letters “tcF”.