Aluminum Cyclone

25 mm - Catalog No. 225-01-01

37 mm - Catalog No. 225-01-02

The SKC Aluminum Cyclone

The SKC Aluminum Cyclone is a lightweight respirable dust sampler that is used with a
fi lter loaded into a three-piece fi lter cassett e. The fi lter material, pore size, and the support
pad must be selected as specifi ed by the sampling method used. The cyclone separates the
dust particles according to size. The respirable particles collect on a fi lter for analysis while
larger particles fall into the grit pot and are discarded.

Flow rate:

2.5 L/min to conform to the ISO/CEN particle size selection criteria and

OSHA proposed rule for silica (4.0-μm 50% cut-point)

*

2.8 L/min for alternate applications (3.5-μm 50% cut-point)

* As previously published, a fl ow rate of 2.6 L/min will give a 4 μm cut-point, however, 2.5 L/min

will give a bett er match over the entire curve.

† Determined using experimental data obtained at fl ows from 2.0 to 4.0 L/min

Note: The ACGIH/ISO/CEN curve requires using the fl ow rate that minimizes the bias of the
cyclone separation from the convention curve.

Operation

Assembly
1. Disassemble a three-piece cassett e and set
aside inlet section (usually marked “inlet”).
Remove plug from cassett e outlet.

2. Select

a

fi lter and support pad as specifi ed

in the sampling method. Insert cassett e ring
(middle) section into cassett e outlet (see Figure 1).

3. Insert cyclone into cassett e ring and press
fi rmly. Leave grit pot in place during calibration
and sampling.

4. Insert

cyclone/cassett e assembly through

large opening of fi lter cassett e holder (see Figure
3
). Ensure assembly is seated fi rmly in holder
by inserting the cyclone‘s side pin (small round
nodule on rim of cyclone) into the notch in the
fi lter cassett e holder.

5. Secure assembly in holder by stretching the
spring-loaded hold-down plate over the cassett e.
Insert adapter (located in end of rubber tubing)
into cassett e outlet.

Calibration
1. Prepare a cyclone/fi lter cassett e assembly (see Assembly).
Calibrate the cyclone with grit pot in place.

2. Moisten O-ring inside large end of calibration chamber (Cat. No. 225-01-
03, see Figure 2) and push chamber over cyclone stem until it fi ts snugly.

3. Connect the Tygon

®

tubing att ached to the fi lter cassett e holder to the inlet

of a sampling pump. Using fl exible tubing, connect the calibration chamber to
a calibrator.

4. Following instructions in the pump and calibrator manuals, calibrate to
required fl ow rate (see reverse side).

5. Disconnect calibrator and replace fi lter used to set fl ow with a fresh
weighed fi lter (unused) for sample collection.

Sampling
1. Ensure

fl ow rate has been

calibrated (see Calibration).

Warning: Leave grit pot in place
during calibration and sampling.

2. Connect the Tygon tubing att ached
to the fi lter cassett e holder to the inlet of
a sampling pump.

3. Clip

cyclone/fi lter cassett e assembly

onto a worker‘s collar or pocket near the
breathing zone.

4. Start pump and record pertinent
details.

5. At the end of the sampling period,
stop pump and record fi nish time.
Remove cyclone/fi lter cassett e assembly from cassett e holder. Separate cyclone from
cassett e. Reassemble inlet section of cassett e, ensuring the inlet and outlet are sealed with
plugs. Send cassett e and all data to a laboratory for analysis.
Note: Particles collected in the grit pot do not represent any part of the respirable dust sample and
should be discarded.

6. Prior to further sampling, clean all parts of the cyclone, including the interior of the grit
pot, with mild soapy water. The cyclone may be wiped with a clean dust-free tissue, air
dried, blown dry, or wiped with isopropyl alcohol.

Accessories

Catalog No.

Three-piece Filter Cassette Blanks

25 mm
37 mm

225-3-25LF
225-3LF

Filter Cassette Holder

25 or 37 mm

225-1

Calibration Chamber

25 or 37 mm

225-01-03

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which provides
SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited
Warranty and Return Policy, go to htt p://www.skcinc.com/warranty.asp.

Form 3803 Rev 1402

Figure 1

Exploded view

of the cyclone/fi lter

cassette assembly

Figure 3

Cyclone with fi lter cassette in holder

Tygon
Tubing

Rubber
Tubing

Cyclone/Filter
Cassette Assembly

Filter
Cassette
Holder

Grit Pot

Collar Clip

Adapter

Air Flow

Figure 2

Calibration

Chamber

For more information: See SKC Technical Note, Publication 1519, at www.skcinc.com.