SKC 225-9019 Preloaded Coated Filter User Manual

SKC Hardware

Advertising
background image

Preloaded Coated Filters

Cat. No. 225-9019

SKC coated fi lters are shipped preloaded in cassett es with end plugs and shrink
bands in place. No assembly is required; they are ready to use.

Method & Chemical: OSHA 81 for crotonaldehyde

Filter & Coating: 2 glass fi ber fi lters treated with 2,4-dinitrophenylhydrazine and
phosphoric acid. Three-piece 37-mm cassett e. Open-face sampling confi guration

Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 3 months.

Sample Stability: Store samples in the dark at -4 F (-20 C). Freezer storage
and shipping required

For information on other available

coated fi lters, request SKC Publication 1219.

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330

Tel: 724-941-9701 Fax: 724-941-1369

www.skcinc.com

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.

How to Use SKC Preloaded Coated Filters

1. Select one coated fi lter cassett e for calibrating the fl ow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the
outlet. Set up the sampling train for calibration with the representative fi lter
cassett e in line. For “closed-face” sampling, remove the plugs and connect the
cassett e to the sampling train. For “open-face” sampling, remove the outlet
plug and the cassett e inlet section then connect the cassett e to the sampling
train. For details on sett ing up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).

2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).

3. Replace

the

cassett e used for calibration with a fresh coated fi lter cassett e

for sample collection. Att ach the cassett e to a worker’s collar, as close to the
breathing zone as possible, by using a fi lter cassett e holder (SKC Cat. No.
225-1). Sample for the specifi ed time interval and record the time.

4. Remove

the

fi lter cassett e at the end of the specifi ed sampling period and

replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using
the same cassett e and fl owmeter used for calibration in Step 2 to ensure that
the fl ow rate has not changed by more than 5%.

5. Store samples at -4 F (-20 C). Appropriately package samples, blanks, and
all pertinent data to send to a lab for analysis. Expedite shipping of samples
to laboratory. Friday shipments are not recommended.

Form 40048 Rev 1303

Advertising