SKC Limited Calibration Adaptor for Cowled Asbestos Sampling Head instructions User Manual

Calidaptor” for cowled asbestos head, O ring seal calidaptor bowl o ring seal hosetail

Advertising
background image

Page 1

007-09-011 Issue B August 2009

www.skcltd.com

“Calidaptor” for Cowled Asbestos Head

Operating Instructions

391-05 “Calidaptor” Calibration
Adaptor
This Calidaptor is designed for use
with the following sample heads:

225-54 Asbestos sampling head
25mm with plastic cowl

225-54A Asbestos sampling
head 25mm with aluminium cowl

A Calidaptor designed for use with
the I.O.M. inhalable, seven hole and
single hole heads is also available
(part no. 391-01).
The SKC Calidaptor allows the flow
setting of a sampling train without
the problems of holding the head,
pump and adjusting the flow all at
the same time.
The Calidaptor incorporates an
‘O’ ring seal which secures the
sampling head in place whilst also
providing a leak proof joint.
The Calidaptor can either be
screwed directly into the outlet of
SKC rotameters (390, 391 and
392 series only) with another ‘O’
ring providing a leak proof joint, or
connected to other flow measuring
devices using the supplied screw-
in hosetail and a length of flexible
tubing.

O ring seal

Calidaptor
bowl

O ring
seal

Hosetail

Advertising