Voltage test overvoltage test current test – Agilent Technologies 6625A User Manual

Page 38

Advertising
background image

Getting Started

38

Voltage Test

1. Set the voltage of the selected output to 10 V by pressing:

VSET

1

0

ENTER


2. Check that the display reads approximately 10 V and 0 A and the CV annunciator is on indicating that the supply is in
the constant voltage mode of operation.


Overvoltage Test

1. Program the overvoltage protection (OVP) to 19 V by pressing:

OV

1

9

ENTER

SET


2. Set the voltage to 16 V by pressing:

VSET

1

6

ENTER


3. Check that the display reads approximately 16 V and 0 A.


4. Set the voltage to 20 V by pressing:

VSET

2

0

ENTER


5. Check that the display reads "OVERVOLTAGE’’


6. Reset the supply by pressing:

VSET

1

6

ENTER

RESET


7. Check that the display reads approximately 16 V and 0 A.


Current Test

1. Turn off the supply.


2. Remove the barrier block cover from the output to be tested and connect a short circuit (jumper wire) between the + V
and -V terminals of the output being tested.


3. Turn on the supply.

4. Use the

to select the output being tested.


5. Set the voltage to a volts by pressing:

VSET

5

ENTER


Set the current to 0.01 amps by pressing:

ISET

.

0

1

ENTER


Advertising