Series IN 500 Operating Manual

Data format UPP (Universal Pyrometer Protocol) 32

Description

Command

Parameters

Reading measuring

value:

AAms

Output: YYYYY; 5-digit decimal, in

1

/

10

°C or °F

88880 = temperature-overflow
75550 = exceed max. sensor head temp.
74440 = fall below min. sensor head temp.
02563 = 256.3 °C or °F; -0170 = -17.0 °C or °F

repeatedly reading
measuring values:

AAmsXXX

XXX = 000 ... 999
XXX = Number of measuring values

Emissivity:

AAemXXXX

XXXX = 0100 ... 1200 in ‰

Response time t

90

:

AAezX

X = 0 ... 6 (decimal)

0 = intrinsic time constant of the device

1 = 0.5 s 2 = 1 s

3 = 2 s

4 = 5 s

5 = 10 s

6 = 30 s

Setting clear time t

CL

:

AAlzX

X = 0 ... 8 (decimal)

0 = Max. value- / min. value storage off

1 = 0.10 s 5 = 5.00 s
2 = 0.25 s 6 = 25.00 s
3 = 0.50 s 7 = external clearing
4 = 1.00 s 8 = automatic clearing

External clearing:

AAlx

Simulation of an external reset contact

Setting maximum /
minimum value:

AAmiX

X = desired setting
0 = maximum value, 1 = minimum value

Read. max/min value:

AAmi

Output: 0 or 1; 0 = max value, 1 = min value

Reading basic
measuring range:

AAmb

Output: YYYYZZZZ (8-digit hex)

2)

YYYY = beginning of measuring range

ZZZZ = end of measuring range

Reading sub range:

AAme

as with mb

Setting sub range:

AAm1YYYYZZZZ

YYYY = beginning of measuring range

ZZZZ = end of measuring range

Analog output

1)

:

AAasX

X = 0 ... 4

2 = 0 ... 5 V

0 = 0 ... 20 mA

3 = thermocouple

type K
1 = 4 ... 20 mA

4 = thermocouple

type J

Temperature display

1)

: AAfhX

X = 0;1 0 = Output Celsius

1

=

Output

Fahrenheit

Reading ambient
temperature:

AAut

Output: stored value, 4-digit hex

2)

e.g. 0258 corresponds to 600 degrees

Entering ambient
temperature:

AAutXXXX

XXXX = Value of ambient temp., 4-digit hex

2)

XXXX z.B. FFEC corresponds to -20 degrees

-99

dez

= FF9D

hex

means: automatic, no

manual compensation

Sensor head temp.:

AAgt

Output: XXX (dec. 000...180 °C or 032...356 °F)

Max. sensor head
temp.:

AAtm

Output: XXX (dec. 000...180 °C or 032...356 °F)

Baud rate:

AAbrX

X = 0 ... 4

0: 1200 bd 1: 2400 bd

2: 4800 bd 3: 9600 bd 4: 19200 bd

Instrument address

1)

:

AAgaXX

XX: device address decimal; 00 ... 31 variable

device addresses; global: 98, 99

Reading switch point:

AAsl

Output: XXXX (4-digit hex)

2)

Setting switch point:

AAslXXXX

XXXX (4-digit hex)

2)

; adjustable within sub

range