Seca 460 RS232 User Manual

Description of new rs232 interface, index c

Advertising
background image

30-34-00-605 Ind. c

1(2)

Rev.: 18.10.05

Description of New RS232
Interface, Index c:


Description of Hardware:

The seca weighing scales Controller owns an
integrated UART-interface which may be configured
by software. The UART is loaded with serial data
out via software. The serial data are transmitted as
3V or 5V leveled signals to a RS232 signal level
converter (e.g. MAX 232). After this conversion
serial data is applicable at the serial interface as
"TXD inverted" and "TXD-GND" for two line serial
data transmission.

Transmission Parameter:

The serial interface is configured as follows:
Baudrate:

1200

Parity:

even

Databits

8

Stopbits

1

Startbits

1

No hardware handshake

RS232 Protocol:

Every second a complete data block of ASCII coded
bytes is transmitted with high nibble first and low
nibble last. Every data block starts with STX and
ends with CR, LF, ETX.

For the kind of data A0 and A1 every data block
owns the following structure with the new protocol:

Byte Name

Description

1

STX

Start of Text

2

KK_0

Identifier new Protocol

KK_1

kind of Data

3

BCD_0

BCD-Information MSDigit

BCD_1

BCD-Information

4

BCD_2

BCD-Information

BCD_3

BCD-Information

5

BCD_4

BCD-Information LSDigit

DISPTYPE

displayed format

6

CTRL_1

control byte 1 - high nibble

CTRL_1

control byte 1 - low nibble

7

CTRL_2

control byte 2 - high nibble

CTRL_2

control byte 2 - low nibble

8

CRC

CRC high byte

CRC

CRC low byte

9

CR

Carriage Return

10

LF

Line feed

11

ETX

End of Text



For the kind of data A4 every data block owns the
following structure with the new protocol:

Byte Name

Description

1

STX

Start of Text

2

KK_0

Identifier new Protocol

KK_1

kind of Data

3

BCD_0

BCD-Information MSDigit

BCD_1

BCD-Information

4

BCD_2

BCD-Information

BCD_3

BCD-Information

5

DISPTYPE

displayed format - high nibble

DISPTYPE

displayed format - low nibble

6

CTRL_1

control byte 1 - high nibble

CTRL_1

control byte 1 - low nibble

7

CRC

CRC high byte

CRC

CRC low byte

8

CR

Carriage Return

9

LF

Line feed

10

ETX

End of Text


KK_0 - Identifier new protocol:
"A": With the new protocol every data block

starts after STX (start of text) with the
identifier indicating you are receiving the
new protocol. It's a ASCII "A"

KK_1 - kind of data
"0": periodic normal weight
"1": single measured value occurred because

of trigger event e.g. measured value
became stable

"4": periodic normal length


REMARK: The amount of "kind of data" may be

broadened for future purposes


BCD_0..BCD_4 - BCD Information

contains the actual binary coded decimals
for each digit without datapoints

DISPTYPE - Displayed format ( for "A0", "A1")
"0": adult weighing scale, kg.g

kkk.gg

"1": adult weighing scale, kg.g

kkk.g

"2": baby weighing scale, kg.g

kk.ggg

"3": adult weighing scale, lbs.lbs

ppp.p

"4": adult weighing scale, lbs

ppp

"5": adult weighing scale, lbs:oz

ppp:oo

"6": adult weighing scale, stones

ss:pp.p

"7": baby weighing scale, lbs:oz

pp.oo.o

"8": by weighing scale, lbs:oz

pp.oo

"9": adult weighing scale, lbs.lbs

ppp.pp

DISPTYPE - Displayed format ( for "A4")
"01": length in cm , unit 1 mm
"02": length in foot-inch
"03": length in inch , unit 0.01 inch















CTRL_1..CTRL_2 - Control byte (for “A0“,“A1“)

control byte 1