5 7 8 m, 999m, 400 series – HT instruments SPEED418 User Manual

Page 24: En - 23, Description of anomalous results

Advertising
background image

400

Series

EN - 23

9.

In case the TMR mode has been selected, pressing the GO/STOP key on
the instrument or the START key on remote probe stops the test before the
set time has elapsed.

10. Should the measured value

be higher than the set limit,
the instrument gives a
double acoustic signal and
displays the message “OK”
which signals the positive
result of the test and a
screen similar to the one
reported here to the side

M

5 7 8 M

5 2 6 V

1 5 s

Insulation

resistance

Applied test voltage and test
duration time

OK

MAN 500V 0.50M 15s

Func

VNom Lim Temp

11. Should the measured value

be higher than the full scale
(§ 12.1), the instrument
gives a double acoustic
signal and displays the
message “OK” which signals
the positive result of the test
and a screen similar to the
one reported here to the side

M

> 999M

5 2 6 V

2 s

Insulation

resistance

Applied test voltage and test
duration time

OK

MAN 500V 0.50M

Func

VNom Lim

12.

The results displayed can be saved by pressing the SAVE key twice or the
SAVE

key and, subsequently, the ENTER key (§ 9.1)


6.3.1. Description of anomalous results

1. Should the instrument not be

able to generate the nominal
voltage, at the end of the
test, it gives a long acoustic
signal and displays and a
screen similar to the one
reported here to the side

M

0.01M

6 4 V

6 s

Not correct

MAN 500V 0.50M

Func

VNom Lim

2. If the measured insulation

value is lower than the set
limit, the instrument displays
a screen similar to the one
reported here to the side and
gives a long acoustic signal

M

0.19M

5 2 6 V

2 s

Not correct

MAN 500V 0.50M

Func

VNom Lim

Advertising
This manual is related to the following products: