9. By

using

the P-PE mode, if

the instrument detects that,
in case the test should be
performed, a contact voltage
higher than the set limit
would build up in the system
being tested, it does not
perform the test and displays
the message reported here
to the side. Check the

L O O P

----

- - - A

FRQ=50.0Hz
VP-N=231V

VP-Pe=231V

Ut > Ulim

Dangerous contact voltage

P-PE STD

Func

Mod.

efficiency of the protective conductor and the earthing

10. If, after repeated tests, the

instrument has overheated,
the message reported here
to the side is displayed. Wait
for this message to
disappear before performing
other tests

L O O P

----

- - - A

FRQ=50.0Hz
VP-N=231V

VP-Pe=231V

Hot temperature

Overheated instrument

P-N STD

Func

Mod.

11.