4.2.2. Special cases which may occur during the tests "MAN", "AUTO", "TIMER"

In case a value of

R

ISO

higher than

R

MAX

was detected (depending on the

selected voltage see following Note
table, the instrument emits a

double

sound signal at the end of the test
indicating the positive outcome of the
test
and displays one screen similar to
the screen alongside.

R

ISO

> 999

M

500 V 20

s


MAN

In case a test is effected at a

voltage

lower than the set rated voltage, at
the end of the test the instrument emits
a

long sound signal and displays one

screen similar to the screen alongside.

R

ISO

o0.01

M

500 V 20

s


MAN


SAVING:

The test can be stored pressing the

SAVE key twice

(according to paragraph 5.1).

If the terminal voltage is higher than

30V, the instrument does not effect the
test, displays the screen alongside for 5
seconds after which, it shows the
default screen relative to the R

ISO

test

under selected mode.

R

ISO

o 33V

VOL AG


THIS RESULTS CANNOT BE SAVED.

ATTENTION: the
test was not
carried out.
Check that the
circuit is not
energised.

The symbol

">"

means that the
resistance value
R

ISO

is higher than

R

MAX

Test duration

Selected mode
MAN

Selected mode
MAN

ATTENTION: the test of resistance R

ISO

was

effected at a voltage value lower than the set
rated voltage
. Low insulation case. This case
occurs under low insulation conditions or in the
presence of capacity on the installation.

Maximum resistance value which can be
measured (999

Ω is displayed if a rated voltage of

500V was selected see table).