O 487, No pe – HT instruments MACROTEST 5035 User Manual

Page 28

Advertising
background image

MACROTEST

5035

EN - 27

If the instrument detects that the earth

cable (green) is not connected, the
screen alongside is displayed for 5
seconds then the initial display is back.
Check the connections of PE conductor
under test.

RCD

~

o

>

1999




no

PE


THE PREVIOUS RESULT CANNOT BE SAVED.


In the tests MAN x1, x2, x5 and AUTO

(during x1 and x5 tests), if the RCD
trips to separate the circuit within a time
not complying with the limits reported in
table, the instrument emits

a long

sound signal at the end of the test and
displays the values alongside.

RCD

~

o 487

ms

Ι

N

30

m

A

1

V

x

1

MAN

t

U

L

=

50

V


If the RCD trips within a time higher

than the instrument’s measuring limits,
the instrument emits

a long sound

signal at the end of the test and
displays the values alongside.

RCD

~

o

>

999

mS

Ι

N

30

m

A

2

V

x

1

MAN

t

U

L

=

50 V


The maximum duration depends on the test type:

Test type

General RCD

Selective RCD

MAN x1 test

999ms

999ms

MAN x2 test

200ms

250ms

MAN x5 test

50ms

160ms

“ ” test

300ms


SAVING:

The tests can be stored pressing the

SAVE key twice

(according to paragraph 5.1).

RCD tripping time

ATTENTION: the
tripping time is
higher than the
standard limit.

RCD tripping
time:
> maximum
measuring time
(it depends on
type of test, see
following table)
.

Message
“no PE”: the
instrument does
not detect the
protection circuit.

SAVE

or RCD

or RCD

SAVE