SOLAR300 - SOLAR300N

EN - 115

This screen (Fig. 185) contemporarily shows the waveforms of phase 1 voltage and current,
scaled according to the Full scale (Fs), and the relevant RMS values; these values have
been saved by the instrument upon pressing the SAVE key.

Fig. 185: Voltage and current scope in single-phase system

The following keys are active on this page:
 The F2 key (or the ZOOM V or ZOOM I item on the dislay) switches between voltage

zoom and current zoom (only if the Manual zoom mode has been activated)

 The up (ZOOM+) or down (ZOOM-) arrow keys zoom in or out the displayed

waveforms (only if the manual zoom mode has been activated)

 The F1 key (or the PAG item on the display) advances to the following page of saved

current values

 The ESC key (or the smart icon on the display) to exit the function and go back to the

“Recording analysis” page (Fig. 181)


This screen (Fig. 186) contemporarily shows the instant values of the waveforms of currents
I1, I2, I3, In scaled according to the Full scale (Fs) and the relevant RMS values; these
values have been saved by the instrument upon pressing the SAVE key.

Fig. 186: Current scope in three-phase 4-wire system

The following keys are active on this page:
 The up (ZOOM+) or down (ZOOM-) arrow keys zoom in or out the displayed

waveforms (only if the manual zoom mode has been activated)

 The F1 key (or the PAG item on the display) advances to the following page of saved

values relative to phase 1 voltage and current

 The ESC key (or the smart icon on the display) to exit the function and go back to the

“Recording analysis” page (Fig. 181)