Caution – HT instruments SOLAR300N User Manual

Page 158

Advertising
background image

SOLAR300 - SOLAR300N

EN - 156

CAUTION

 When SOLAR300N is set in order to use MPP300 as a remote unit, ALL

connections relevant to electrical quantities (voltages and currents) must
be carried out on unit MPP300. SOLAR300N must have no voltage nor
current
connected to its inputs.

 The maximum voltage for the inputs of MPP300 is 1000VDC between

inputs VDC1, VDC2, VDC3 and 600VAC between inputs VAC1, VAC2,
VAC3. Do not measure voltages exceeding the limits given in this manual.
Exceeding these limits could result in electrical shocks to the user and
damage to the instrument.

 In order to guarantee the operator’s safety, while making the connections,

disable the system being measured by means of the switches/breakers
upstream and downstream of the DC/AC converter (inverter).


1. Check and, if necessary, set the sensitivity of the reference cell on SOLAR-02 consistently

with the type of PV modules which will be measured (please refer to the User Manual of
SOLAR-02).

2. We recommend making a preliminary evaluation of the value of irradiation on the surface of

the PV modules being tested by means of unit SOLAR-02 (operating independently) and the
reference cell. We suggest carrying out the tests only if irradiation is higher than 600W/m

2

3. Switch on SOLAR300N, check and, if necessary, change the settings relevant to the type of

remote unit, to the minimum irradiation threshold, to the full scale of the AC and DC clamps
and to the parameters of the system being measured (see §5.3.3.1, §5.3.3.3, §5.3.1.2,
§5.3.1.3, §5.3.2).

4. In order to guarantee the operator’s safety, disable the system being measured by means of

the switches/breakers upstream and downstream of the DC/AC converter (inverter).

5. Connect SOLAR300N to the MPP300 unit via the USB cable and bring SOLAR-02 and

MPP300 nearer (max distance approx. 1 m). All instruments must be switched on (see
the User Manuals of SOLAR-2 and MPP300 for further details). The display of SOLAR300N
must show (for 5 seconds) the message “MPP300 connected”.

6. On SOLAR300N, access the GENERAL MENU, select the function View Measures and

press ENTER; wait for the three units to start communicating with each other. This condition
is highlighted by the contemporary presence of the following indicators:

a. Symbol

steady (not flashing) on the display of SOLAR-02

b. MASTER and REMOTE LEDs flashing green on unit MPP300

7. Fig. 246Fig. 247Connect the VDC1(+) and VDC1(-) inputs of unit MPP300 to the output

terminals of the string, respecting the polarities and the colours indicated in Fig. 246 or
Fig. 247.

8. Repeat the operation described in the step above for other possible DC power trackers to be

monitored by using the VDC2 and VDC3 inputs according to the number of DC inputs set
(see § 5.3.2.5).

9. Connect the output connector of the DC clamp to the IDC1 input of unit MPP300.


Advertising
This manual is related to the following products: