Caution – HT instruments SOLAR300N User Manual

Page 7

Advertising
background image

SOLAR300 - SOLAR300N

EN - 5

1.3. DURING

USE

Please read carefully the below points:

CAUTION

 Should you fail to keep to the prescribed instructions you could damage

the instrument and/or its components or endanger your safety

 Do not touch any unused terminal when the meter is connected to the

circuit on test

 When measuring current, other currents located near the leads may affect

the measuring accuracy

 When measuring current, always position the wire in the very middle of the

jaws in order to obtain the highest accuracy

 A measured value remains constant if the "HOLD" function is active.

Should you notice that the measured value remains unchanged, disable
the “HOLD” function


1.4. AFTER

USE

 After use, turn off the instrument by pressing ON/OFF key for a few seconds.

 If you expect not to use the instrument for a long time please keep to the storage

instructions described at § 3.4.


1.5. MEASURING

(OVERVOLTAGE) CATEGORIES DEFINITIONS

IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control
and laboratory use, Part 1: General requirements, gives a definition of measuring
category, usually called overvoltage category. § 6.7.4: Measuring circuits:

(OMISSIS)

Circuits are divided into the following measurement categories:

Measurement category IV is for measurements performed at the source of the low-

voltage installation

Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units

Measurement category III is for measurements performed in the building installation.

Examples are measurements on distribution boards, circuit breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary
motors with permanent connection to fixed installation

Measurement category II is for measurements performed on circuits directly

connected to the low voltage installation

Examples are measurements on household appliances, portable tools and similar
equipment

Measurement category I is for measurements performed on circuits not directly

connected to MAINS

Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of the
equipment is made known to the user