Caution – HT instruments SOLAR300N User Manual

Page 93

Advertising
background image

SOLAR300 - SOLAR300N

EN - 91

Press GO/STOP key again to start recording and verify the possible residual error on
window message.
Confirm with ENTER or press “Ok” or “Cancel” keys to close the warning window message
and start the recording by pressing GO/STOP key anyway.

As the default value of the integration periods is set at 15min (see § 10.11.1) the meter will
store data in the temporary memory for this time. Afterwards the meter will process the
results saved in the temporary memory and will save the first series of values in the
definitive memory. Therefore, if an integration period of 15 minutes has been set, the
recording will continue for about 15 minutes before producing a series of recorded values

CAUTION

Let the meter record for at least one integration period in order to save a
valid value. If the recording is interrupted before the selected integration
period has completely elapsed the data stored in the temporary memory will
not be processed and the corresponding series of values won’t be
transferred to the definitive memory.


5.6.1. Recording settings screen – NPV systems (only for NPV systems)
To start a recording under automatic mode a preliminary definition of a correct starting
Date/Hour, compatible with system date, is needed inside “Recorder Settings” section (see
§ 5.4). The below screen is shown:

12/09/2006 – 16:55:10

RECORDING SETTINGS

Comments: POWER & ENERGY

Start:Auto 12/09/2006 – 16:30

Stop:Manu

Integration Period:15 min

Cogeneration

+ General

Parameters

Start:

Auto

12 / 09 / 06 – 17 : 00 : 00

MOD(+)

MOD(-)

Fig. 147: Correct Date/Hour

12/09/2006 – 16:55:10

RECORDING SETTINGS

Comments: POWER & ENERGY

Start:Auto 12/09/2006 – 16:30

Stop:Manu

Integration Period:15 min

Cogeneration

+

General

Parameters

Error: wrong start date

CMP/EXP

PREDEF.

SEL

MODIFY

Fig. 148: Wrong Date/Hour


1. Move the cursor using arrows keys on field relative to “Start” (and or the “Stop” ),

marked them with blue background

2. Press F4 key (alternatively touch the MODIFY at display). In the bottom of display a

command bar with “Manu” is shown. Press F3 key (MOD(+)) or F4(MOD(-)) and select
Auto

3. Use left or right arrows keys to move on date and hour fields. Using the up arrow key or

F3 (MOD(+)) key to increase the value and the down arrow key or F4 (MOD(-)) key to
decrease the value. Set a correct Date/Hour as shown in Fig. 147

4. Press SAVE or ENTER keys (or smart icon

) to saving the settings. The screen of

Fig. 148 is shown if a wrong Date/Hour is set. Under this condition the SAVE key does
not have any effect. Repeat the steps 2 and 3 and set corrects values before pressing
SAVE key again

Advertising
This manual is related to the following products: