HT instruments VEGA78 User Manual

Page 10

Advertising
background image

VEGA78

EN - 8

4.2. KEYBOARD

DESCRIPTION

The below keys are available:

 Key

ON/OFF:

Press this key to turn on the meter. Press and hold the key for few
seconds to turn off the meter.

 Keys

F1, F2, F3, F4: Multifunction keys. The various functions are intended from the s

shown on the bottom of display.

 Key

ESC:

To leave a menu or a sub-menu. The smart icon which is shown
on screens performs the same function under interactive mode.

 Key

ENTER:

Double function key:

ENTER:

To confirm the settings made.

HOLD:

To block the value updating in all real time screens and also
during a recording operation. “H” symbol is shown at display by
any pressing of key as for Fig. 4:

12/09/2006 – 16:55:10

H

TOTAL RMS VALUES – Page 1/6

V1N

0.0

V2N

0.0

V3N

0.0

VNPE

0.0

V

V12

0.0

V23

0.0

V31

0.0

V

NEG%

0.0

ZERO%

0.0

SEQ

000

Hz

0.0

I1

0.0

I2

0.0

I3

0.0

IN

0.0

A

PAGE

SCOPE

HARM

VECTORS

Fig. 4: HOLD function active at display


 Keys

, , , /ZOOM in/ZOOM out: These arrows keys permit to select the desired

programmable parameters inside various screens. The double
function of ZOOM in and ZOOM out permits to change manually
the full scale of some graphics inside SCOPE section in order to
improve the global resolution on viewing waveforms of signals (see
§ 5.3.2). Arrow keys

and  permit to operate on internal pages

inside some kinds of screen.

 Key

SAVE: Saves

an

Instant” type sampling of instantaneous values in

internal memory (see § 5.5). The same key permits also the saving
of various settings inside screens. The smart icon

which is

shown on screens perform the same function under interactive
mode.

 Key

GO/STOP:

Permits to start/stop any recording (see § 5.4.10).

 Key

HELP:

Opens an help on line window on the meter screen with a short
description of the same screen which is instantaneously shown.
This key is active for any function.

Advertising