HT instruments VEGA78 User Manual

Page 49

Advertising
background image

VEGA78

EN - 47

5.4.6.1. General Parameters: sub-levels description

Press F1 key (or touch CLP/EXP at display) to expand or compress the sub-levels. The
parameters inside sub-levels are strictly depending on type of selected system (see §
5.3.1). Below there are some pictures of different possible situations:

12/09/2006 – 16:55:10

RECORDING SETTINGS

- General Parameters

Frequency

- 

Voltages

Phase 1-Neutral

+

Currents

+

Active Power&Energy

8/251 Parameters – Autonomy: 109d 0h

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 91: General Parameters sub-level - Single phase system


Each parameter is always selectable independent from others. The below parameters
can be selected for recording in Single phase systems:

Parameter Description

Frequency

Frequency of phase L1

Voltages

TRMS voltages L1-N and N-PE

Currents

TRMS current phase L1

Active Power & Energy

Active power and energy of phase L1

Reactive Power & Energy

Reactive (inductive or capacitive) power and energy
of phase L1

Apparent Power & Energy

Apparent power and energy of phase L1

Power Factor

Power factor of phase L1

CosPhi

Power factor referred to fundamental of voltage and
current of phase L1

Table 2: Selectable parameters for Single phase system

Press SAVE or ENTER keys (or the smart icon

) to save each selection and confirm by

“Ok”. The main screen of Fig. 85 is shown by meter at the end of the operation.
Press ESC key (or the smart icon ) to quit without saving and back to previous screens.

12/09/2006 – 16:55:10

RECORDING SETTINGS

- General Parameters

Frequency

- 

Voltages

Phase1-Neutral

Phase2-Neutral

Phase3-Neutral

32/250 Parameters – Autonomy: 27d 6h

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 92: General Parameters sub-level - Three phase 4-wire system

Advertising