In-Situ TROLL 9500 Quick Start Guide User Manual

Mp troll 9500, Quick start guide

Advertising
background image

Multi-Parameter

Water Quality

TROLL

®

C

D

6

Connect to the TROLL 9500—

7

When connected, the device appears in the Navigation tree
below the COM port. Select the TROLL and set its clock—

MP

TROLL 9500

Quick STaRT Guide

5

Launch the software—

Win-Situ

: Double-click the desktop shortcut

.

A

Select the TROLL. The software retrieves and displays informa-

tion about the device and all installed sensors. This can take a
moment . . .

B

Select

Edit

C

In the next screen, select the

Clock

option

D

Follow the Wizard to synchronize the device clock with the PC

clock

A

Highlight the

COM

port in the Navigation tree

B

Click or tap

Find

8

Quick-Cal the Basic water quality sensors

A

Detach the Cal Cup. Empty out liquid and remove small sponge (if present). Rinse

sensors with DI water

B

Fill Cal Cup to the line with Quick Cal solution and re-attach—not too tight

C

Establish software connection as above. Select Parameters in the Navigation tree

D

Click or tap

Quick Cal

. Select sensors to calibrate. Click or tap

Run

to start pH,

ORP, and Conductivity calibration

E

For Polarographic DO

: Turn probe upside-down and pour out some solution so

DO sensor membrane is in air and temperature sensor is completely covered. Re-
attach cap loosely. Select vented cable or supply barometric pressure for accurate
results. Select

Run

to complete DO calibration. Details are in Section 3 of the

Operator’s Manual

F

Remove Cal Cup, pour out the Quick Cal solution, rinse the front end of the probe,

and attach the restrictor

E

B

Your new Multi-P arameter TROLL 9500 arrives with the following installed:
• Removable Basic sensors pH, ORP, Polarographic Dissolved Oxygen, Conductivity (as required)
• Alkaline batteries—if shipped with Polarographic DO sensor, to keep the sensor conditioned
• Permanent sensors Barometric Pressure, Temperature, Pressure/Level (optional), Turbidity (optional)
The Cal Cup is attached with water in the bottom to keep the sensors hydrated

A

C

B

Battery

compartment

Sealed

body

Restrictor

Nose cone

TROLL

9500

USB or Serial port

Serial

RuggedReader clock. Set
it first if wrong (Start Menu
> Settings > System tab >
Clock)

A

B

OR

Pocket-Situ

: Tap the Start menu, tap Pocket-Situ 4

B

A

Win-Situ 4

A

B

Pocket-

Situ 4

Mating the Twist-Lock connectors is
easy after you do it once. Detailed
steps and photos are on the back of this
Quick Start Guide.

TIP:

Set up the port with a

Direct

connection

to

One

device at

19200

baud

port

1

Install

Win-Situ

®

4

to your local hard drive from

the In-Situ software CD or the In-Situ website.
Optional: Install the desktop component of Pocket-

Situ 4 to the same computer. Then use the In-Situ Software
Manager to install it on the RuggedReader

®

at your next

ActiveSync

®

connection.

3

Attach

a TROLL Com to the TROLL

9500 or to its cable.

Tip: After plugging in a USB TROLL Com, follow the prompts
to install the drivers. Then use Windows

®

Device Manager to

determine which COM port the USB TROLL Com is using.

2

Install batteries:

Tip: If using a USB TROLL Com, be sure to
select the option “Install USB TROLL Com
Drivers” when installing Win-Situ.

4

Plug the TROLL Com into a USB or serial port
on a desktop/laptop PC, or into the serial port
on a RuggedReader handheld PDA.

A Remove battery compartment cover
B Insert batteries
C Replace cover and tighten

On first connection, be sure to select the
correct COM port for a USB connection.
Set up the port for a Direct connection to

One

device at

19200

baud.

Direct Connect

TROLL

Com

Cable Connect

TROLL

Com

RuggedCabl

e™

Advertising
This manual is related to the following products: