KYORITSU 2012R User Manual

Kew mate 2012r, Digital multimeter with ac/dc clamp sensor

Advertising
background image

DIGITAL MULTIMETER

WITH

AC/DC CLAMP SENSOR

KEW MATE 2012R

This instrument satisfies the marking requirement defined

in the WEEE Directive. This symbol indicates separate

collection for electrical and electronic equipment.
This marking means they shall be sorted out and collected

as ordained in DIRECTIVE 2006/66/EC.

This directive is valid only in the EU. When you remove

batteries from this product and dispose them, discard them

in accordance with domestic law concerning disposal. Take

a right action on waste batteries, because the collection

system in the EU on waste batteries are regulated.

#

WARNING

Measurement Category(CAT)

The restrictions on the maximum voltage level for which

the this product can be used, depend on the measurement

categories specified by the safety standards.

Do not apply any input level higher than maximum allowable

input.

AC/DC 600V CAT II    AC/DC300V CAT III

CAT II

Appliances, portable equipment, ect. For measurements

performed on circuits directly connected to the low

voltage installation.

CAT III D i s t r i b u t i o n b o a rd , c i rc u i t b re a k e r, e c t . F o r

measurements performed in the building installation.

#

DANGER

Never make measurement on circuits with a maximum voltage

difference of 600V or greater between conductors (300V or

greater between a conductor and ground).

Do not attempt to make measurement in the presence of

flammable gasses.

Otherwise, the use of the instrument may cause sparking,

which leads to an explosion.

Never attempt to use the instrument if its surface or your hand

is wet.

Do not exceed the maximum allowable input of measuring

ranges.

Never open the battery compartment cover while making

measurement.

Never try to make measurement if any abnormal conditions,

such as broken Clamp Sensor or case is noted.

The instrument is to be used only in its intended applications or

conditions.

Otherwise, safety functions equipped with the instrument

doesn’t work, and instrument damage or serious personal

injury may be caused.

1SAFETY WARNINGS

This instrument has been designed and tested according to

IEC Publication 61010: Safety Requirements for Electronic

Measuring Apparatus. This instruction manual contains warnings

and safety rules which must be observed by the user to ensure

safe operation of the instrument and to retain it in safe condition.

Therefore, read through these operating instructions before

starting using the instrument.

#

WARNING

Read through and understand instructions contained in this

manual before starting using the instrument.

Save and keep the manual handy to enable quick reference

whenever necessary.

Be sure to use the instrument only in its intended applications

and to follow measurement procedures described in the

manual.

Be sure to understand and follow all safety instructions

contained in the manual.

Failure to follow the above instructions may cause injury, damage

to the instrument and/or damage to equipment under test.

The symbol # indicated on the instrument means that the user

must refer to related parts of the manual for safe operation of

the instrument. Be sure to carefully read the instructions following

each # symbol in this manual.

#

DANGER:is reserved for conditions and actions that are

likely to cause serious or fatal injury.

#

WARNING:is reserved for conditions and actions that can

cause serious or fatal injury.

#

CAUTION:is reserved for conditions and actions that can

cause minor injury or instrument damage.

Following symbols are used on the instrument and in the

instruction manual. Attention should be paid to each symbol to

ensure your safety.

#

Refer to the instructions in the manual.

This symbol is marked where the user must refer to the

instruction manual so as not to cause personal injury or

instrument damage.

Indicates an instrument with double or reinforced insulation.
Indicates that this instrument can clamp on bare conductors

when measuring a voltage corresponding to the applicable

Measurement category, which is marked next to this symbol.

Indicates AC (Alternating Current).

Indicates DC (Direct Current).

#

WARNING

Never attempt to make any measurement, if any abnormal

conditions are noted, such as broken case, cracked test leads

or Clamp Sensor Cable and exposed metal parts or internal

wiring.

Do not turn the Function Selector Switch while the test leads

are connected to the circuit under test.

Do not install substitute parts or make any modification to

the instrument. Return the instrument to Kyoritsu or your

distributor for repair or re-calibration.

Do not try to replace the batteries if the surface of the

instrument is wet.

Always disconnect the clamp sensor and the test leads

from the circuit under test and switch off the instrument

before opening the battery compartment cover for battery

replacement.

A cap is provided on the tip of a test lead. Use a test lead with

the cap on for safety.

#

CAUTION

Make sure that the Function Selector Switch is set to an

appropriate position before making measurement.

Always make sure to place the test leads in the holster before

making current measurement.

Do not expose the instrument to the direct sun, extreme

temperatures or dew fall.

This instrument isn't dust & water proofed. Keep away from

dust and water.

Be sure to set the Function Selector Switch to the "OFF"

position after use. When the instrument will not be used for

a long period of time, place it in storage after removing the

batteries.

Use a damp cloth and detergent for cleaning the instrument.

Do not use abrasives or solvents.

3SPECIFICATIONS

Measuring Ranges and Accuracy (at 23 ℃ ±5 ℃ , relative

humidity75% or less)

AC Current  A (RMS value detection)

Maximum Input Current : 120A

Range

Display range

Allowable input

Accuracy

60A

0.00~60.39A

0.00~60.00Arms

(85Apeak or less)

±2.0%rdg±5dgt

(45~65Hz)

sine wave)

120A

0.0~603.9A

0.0~120.0Arms

(170Apeak or less)

For non-sinusoidal waveforms, add ±(2% of reading + 2% of

full scale), for Crest factor<2.5.

DC Current  A

Maximum Input Current : 120A

Range

Display range

Allowable input

Accuracy

60A ±0.00~60.39A

±0.00~60.00A

±2.0%rdg±8dgt

120A ±0.0~603.9A

±0.0~120.0A

±2.0%rdg±5dgt

AC Voltage  V (RMS value detection, Auto-range)

Maximum Input Voltage : 600V

Range

Display range

Allowable input

Accuracy

6V

0.000~6.039V

0.300~600.0Vrms

(850Vpeak or less)

±1.5%rdg±5dgt

(45~400Hz)

sine wave)

60V

5.60~60.39V

600V

56.0~603.9V

Input Impedance:approx. 10MΩ <200pF

For non-sinusoidal waveforms, add ±(2% of reading + 2% of

full scale), for Crest factor<2.5.

DC Voltage  V (Auto-range)

Maximum Input Voltage : 600V

Range

Display range

Allowable input

Accuracy

600mV ±0.0~603.9mV

±0.0m~600.0V

±1.0%rdg±3dgt

6V

±0.560~6.039V

60V

±5.60~60.39V

600V

±56.0~603.9V

Input Impedance:approx. 10MΩ

Resistance Ω (Auto-range)

Range

Display range

Allowable input

Accuracy

600Ω

0.0~603.9Ω

0.0Ω~60.00MΩ

±1.0%rdg±5dgt

6kΩ 0.560~6.039kΩ

60kΩ 5.60~60.39kΩ

600kΩ 56.0~603.9kΩ

6MΩ 0.560~6.039MΩ

±2.0%rdg±5dgt

60MΩ 5.60~60.39MΩ

±3.0%rdg±5dgt

Open-loop Voltage:approx.0.6V, Measuring Current : 0.3mA or

less

Continuity

Range

Display range

Allowable input

Accuracy

600Ω

0.0~603.9Ω

0.0~600.0Ω

±1.0%rdg±5dgt

The buzzer turns on for resistances lower than 35±25Ω.

Open-loop Voltage:approx.0.6V, Measuring Current : 0.3mA

or less

Diode 

Range

Display range

Allowable input

Accuracy

2V

0.000~1.999V

0.000~1.999V

±3.0%rdg±5dgt

Open-loop Voltage:approx.2.7V

Capacitor

(Auto-range)

Range

Display range

Allowable input

Accuracy

40nF

0.00~40.39nF

the accuracy is not guaranteed

400nF

36.0~403.9nF

40.0n~40.00μF ±2.5%rdg±10dgt

4μF

0.360~4.039μF

40μF

3.60~40.39μF

400μF 36.0~403.9μF

the accuracy is not guaranteed

4000μF 360~4039μF

Frequency Hz ( AC Current ) (Auto-range)

Range

Display range

Allowable input

Accuracy

10Hz

0.000~9.999Hz

The accuracy is not guaranteed

100Hz

9.00~99.99Hz

9.00Hz~400.0Hz

±0.2%rdg±2dgt

1000Hz

90.0~400.0Hz

±0.1%rdg±1dgt

400.1~999.9Hz

The accuracy is not guaranteed

10kHz

0.900~9.999kHz

100kHz

9.00~99.99kHz

1000kHz

90.0~999.9kHz

10MHz

0.900~9.999MHz

Input Current:more than 6A

Frequency Hz ( AC Voltage ) (Auto-range)

Range

Display range

Allowable input

Accuracy

10Hz

0.000~9.999Hz

The accuracy is not guaranteed

100Hz

9.00~99.99Hz

9.00Hz~300.0kHz

±0.2%rdg±2dgt

1000Hz

90.0~999.9Hz

±0.1%rdg±1dgt

10kHz

0.900~9.999kHz

100kHz

9.00~99.99kHz

300kHz

90.0~300.0kHz

1000kHz 300.1~999.9kHz

The accuracy is not guaranteed

10MHz

0.900~9.999MHz

Input Voltage:more than 6V (~10kHz) , more than 20V (10k

~300kHz)

Input Impedance:approx. 900kΩ

Note:◇ The symbol of "―" in the above table means that the

instrument only displays the value, but the accuracy, the

proper operation and the safety are not guaranteed.

2FEATURES

Permits AC/DC current measurement up to 120A using a

clamp sensor that comes standard with the instrument

Clamp sensor for ease of use in crowded cable areas and

other tight places

Permits current measurement with an open current-clamp sensor

that does not require opening and closing operations by the user

True-RMS measurements ACV and ACA.

Auto-power-save function

Buzzer for easy continuity checking

Data hold function to freeze the readings

LCD with a bar graph

Shock absorbing holster for ease of storage

Designed to international safety standard IEC61010-1: over-voltage

category CAT III 300V, CAT II 600V and pollution degree 2.

[ Effective Value (RMS) ]

Most alternating currents and voltages are expressed in

effective values, which are also referred to as RMS (Root-Mean-

Square) values.

The effective value is the square root of the average of square

of alternating current or voltage values. Many clamp meters

using a conventional rectifying circuit have "RMS" scales for AC

measurement. The scales are, however, actually calibrated in

terms of the effective value of a sine wave though the clamp

meter is responding to the average value. The calibration is

done with a conversion factor of 1.111 for sine wave, which

is found by dividing the effective value by the average value.

These instruments are therefore in error if the input voltage or

current has some other shape than sine wave.

[ CF (Crest Factor) ]

CF (Crest Factor) is found by dividing the peak value by the

effective value.

Examples: Sine wave: CF=1.414

Square wave with a 1: 9 duty ratio: CF=3

Safety Standard

IEC 61010-1

CAT III 300V, pollution degree 2

CAT II 600V, pollution degree 2

IEC 61010-031

IEC 61010-2-032

IEC 61326(EMC)

Operating System

ΔΣ

modulation

Display

Liquid crystal display

Maximum Reading:6039

Except Hz : 9999, CAP : 4039,

Diode : 1999

Bar graph with maximum points of 30.

Display

approx. 3 times per second

Indication renewal

Location for use

Indoor use, 2000m max, above sealevel

Operating Temperature 0 ~ +40℃ , relative humidity 85% or less

and Humidity Range

(without condensation)

Storage Temperature -20 ~ +60℃ , relative humidity 85% or less

and Humidity Range

(without condensation)

Source

Two 1.5VDC R03 (UM-4) batteries

Current Consumption approx. 3mA (DCV), approx. 13mA

(ACA)

Power-save Function Shifts to the power-save state about

15 minutes after the last switch

operation.

Low battery warning

”Appears when the batteries

become low (2.4±0.15V or less)

Overload Protection

AC voltage / DC voltage / Frequency

ranges :

DC / ACrms 720V for 10 seconds

AC current / DC current ranges :

DC / ACrms 150A for 10 seconds

Resistance / Continuity / Diode /

Capacitor ranges :

DC / ACrms 600V for 10 seconds

Withstand Voltage

AC3540Vrms for 5 seconds between

electrical circuit and housing case

Insulation Resistance 100MΩ or greater at 1000V

between electrical circuit and housing

case

Conductor Size

Approx. φ12mm diameter max

Dimensions

128(L)×92(W)×27(D)mm

Weight

Approx. 220g

Accessories

Two R03 (UM-4) batteries

Instruction Manual

INSTRUCTION MANUAL

Waveform

Crest factor

CF

Average value

Vavg

Effective value

Vrms

Conversion

factor

Vrms/ Vavg

Reading errors for

average sensing

instrument

≒0.637

≒1.111

≒1.414

≒1.155

≒1.732

=-3.8%

×100%

=11.1%

≒0.707

2

1

2

2

π

A

3

D

1

3

2

D

1

3

A

0.5A×1.111-

3

A

A

A

A

A

1

0.5A

1

0%

2 A

π

AD

f

A =A・D

T

2

A

A×1,111-A Ч100

Ч100

D

A

D

1

A

D

A

(1.111  -1)

D

A

0

A

0

A

0

A
0

T

D=f/T

f

3

DISTRIBUTOR

Kyoritsu reserves the rights to change specifications

or designs described in this manual without notice

and without obligations.