Followings are corrections and additional explanations for KEW3128 Instruction manual.

The contents of this sheet apply to the instrument with firmware ver.1.12 or later.



Specification tale for [Insulation Resistance Tester] on page 10 & 11 of the Instruction manual for KEW3128 is
modified as follows. The lower accuracy limit, display range and out-of-range display at each Voltage Range
are added.
[Insulation Resistance Tester]

Rated Voltage

500V

1000V

2500V

5000V

10000V

12000V

Max value

500GΩ

1.00TΩ

2.50TΩ

5.00TΩ

35.0TΩ

35.0TΩ

Accuracy

400k to 50GΩ

±5%rdg±3dgt

800k to 100GΩ

±5%rdg±3dgt

2M to 250GΩ
±5%rdg±3dgt

4M to 500GΩ
±5%rdg±3dgt

8M to 1TΩ
±5%rdg±3dgt

8M to 1TΩ
±5%rdg±3dgt

50G to 500GΩ
±20%rdg
*Accuracy is
not guaranteed
with setting of
250V or less.

100G to 1TΩ
±20%rdg

250G to 2.5TΩ
±20%rdg

500G to 5TΩ
±20%rdg

1T to 10TΩ
±20%rdg

1T to 10TΩ
±20%rdg

10T to 35TΩ
Values are
displayed, but
accuracy isn’t
guaranteed.

10T to 35TΩ
Values are
displayed, but
accuracy isn’t
guaranteed.

Display Range

400k to 999k
1.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 600G

800k to 999k
1.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 999G
1.00T to 1.20T

2.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 999G
1.00T to 3.00T

4.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 999G
1.00T to 6.00T

8.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 999G
1.00T to 9.99T
10.0T to 35.0T

8.00M to 9.99M
10.0M to 99.9M
100M to 999M
1.00G to 9.99G
10.0G to 99.9G
100G to 999G
1.00T to 9.99T
10.0T to 35.0T

Out-of-range
display

Upper limit

<400kΩ

<800kΩ

<2.00MΩ

<4.00MΩ

<8.00MΩ

<8.00MΩ

Lower limit

>600GΩ

>1.20TΩ

>3.00TΩ

>6.00TΩ

>35.0TΩ

>35.0TΩ

Short-circuit current

Max 5.0mA

Leakage current

0.01nA∼

Output current

1mA or more,
1.2mA or less
under a load of
0.5MΩ
*Should be
500V or more

1mA or more,
1.2mA or less
under a load of
1MΩ

1mA or more,
1.2mA or less
under a load of
2.5MΩ

1mA or more,
1.2mA or less
under a load of
5MΩ

0.15mA or
more, 0.25mA
or less under a
load of 10MΩ

0.15mA or
more, 0.25mA
or less under a
load of 12MΩ

Note: The Lower limit value within the Display Range is shown when the test leads are short-circuited and the

Upper limit value within the Display Range is shown when measured values exceeds the Display Range.



P12 Additional explanation to [Ammeter]
[Ammeter]

Measuring Range

0.00nA∼ 2.40mA

Minimum resolution 0.01nA

(determined by resistance and voltage values)

Depending on the effective range