Model 8112 Instruction Manual

Model 8112 clamp adaptor is designed to be an AC current / voltage conversion probe capable of measuring

AC current from 0.1mA to 80A in conjunction with high sensitivity AC voltmeters.

It is especially useful for digital multimeters.

Simple to operate, the unit has range switch position at 200mA, 2A and 20A.
It also features safety design throughout with no exposed metal parts.

Specifications:

Range Switch

Position

Input

Output

Accuracy

Frequency

Range

0 – 0.5A AC

0 – 500mV AC

±1.5% of output ±0.2mV

50Hz–1kHz

200mA / 200mV

(1mA / mV)

0 – 1A AC

0 – 1000mV AC

±3.0% of output ±0.4mV

40Hz–10kHz

±1.0% of output ±0.1mV

40Hz–1kHz

2A / 200mV

(10mA / mV)

0 – 20A AC

0 – 2000mV AC

±1.5% of output ±0.2mV

1kHz–10kHz

±1.0% of output ±0.1mV

40Hz–1kHz

0 – 20A AC

0 – 200mV AC

±1.5% of output ±0.2mV

1kHz–10kHz

20 – 40A AC

200mV – 400mV AC

±2.5% of output

50Hz–10kHz

20A / 200mV

(100mA / mV)

40 – 80A AC

400mV – 800mV AC

±2.5% of output

100Hz–10kHz

Output Impedance – 2k ohms

Frequency Characteristics – Within ±3dB at 30Hz to 100kHz

Maximum Voltage of Circuit under test – 500V (DC + AC peak when transformer jaws closed)

Output Cord – Insulated, approx. 1.5m long, 3.9mm diameter and terminated with banana plug

with barrier

.


Operating Instructions:

After connecting Model 8112 to the voltmeter pull the jaw lever down to open the transformer jaws and

clamp onto a current carrying conductor.
The voltmeter reads the current measured.

Change the range switch position on Model 8112 or the voltmeter according to the current being measured.


Read the current as follows:

Model 8112 set at: AC Voltmeter set at: How to read:


200mA 200mV (e.g.) Read directly (in milliamps)

2A 200mV (e.g.) Divide reading by 100 (read in amps)


20A 200mV (e.g.) Divide reading by 10 (read in amps)