ROTRONIC HC2-S3C03 User Manual

Hc2-s3c0x, Series, Short instruction manual

Advertising
background image

HC2-S3C0X Meteorological Probe
HC2-S3C0X-PTxx Meteorological Probe with passive Pt100

General description

Your ROTRONIC probe is designed according to ISO 9001 and meeting EN61000-6-4 and
EN61000-6-2 EMC requirements. The probe works perfectly for many years when you follow
the operating guidelines.

These short instructions are limited to a description of the main functions and installation of
the device. The detailed instruction manual can be found on the internet at:
www.rotronic-humidity.com

HC2-S3C0X

HC2-S3C0X-PTxx

Cable length in meter

Pt type

Technical Data
Operating & measuring range:

0...100 %rh / -50…100 °C

Outputs (factory default):

0… 1 V = -40...60 °C / 0…100 %rh

Digital output:

UART

Start-up time / Measurement interval: 1.5 s / 1 s
Supply voltage / Load:

5…24 VDC / 5…16 VAC / > 1 kOhm

Accuracy at 23 °C:

HC2-S3C0X:

±1 %rh / ±0.2 K

HC2-S3C0X-PTxx: ±1 %rh / ±0.1 K (passive)

Protection rating:

IP65 (except sensor / fi lter area)

HC2-S3C0X-

SERIES

SHORT INSTRUCTION MANUAL

Wire colours HC2-S3C0X

HC2-S3C0X-PTxx

HC2-S3C0X-PTxx

Green

Power +, max 24 VDC

Power +, max 24 VDC

Pt100 wiring

Gray

GND (digital and power)

GND (digital and power)

Red

RxD (UART)

RxD (UART)

Blue

TxD (UART)

TxD (UART)

White

Analogue signal %rh

Analogue signal %rh

Brown

Analogue signal °C

Analogue signal °C

Shield

Not connected on probe side

AGND (analogue ground)

Black

Ptxx

Purple

Ptxx

Yellow

AGND (analogue ground)

Ptxx

Orange

Ptxx

Calibration and Adjustment
The probe has been factory-adjusted and comes with a calibration certifi cate. We recommend
calibration after about 1 year, depending on use. For probe adjustment instructions, contact
ROTRONIC or the ROTRONIC representative in your country.

Dimensional Diagram

The electronics of the probe is sensitive to electrostatic discharge (ESD) and wrong supply
polarity. Make sure to be properly earthed when handling the probe. Improper handling and
too high supply voltage as well as wrong connections (supply voltage on signal wires) may
destroy the probe. This can be traced and will void warranty!

Accessories for programming and calibration the device
AC3001-XX USB cable with open end to connect the probe to a PC
AC3002-XX RS-232 cable with open end to connect the probe to a PC

ROTRONIC AG, CH-8303 Bassersdorf, Tel. +41 44 838 11 44, www.rotronic-humidity.com
ROTRONIC Messgeräte GmbH, D-76275 Ettlingen, Tel. +49 7243 383 250, www.rotronic.de
ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg, Tél. +33 1 60 95 07 10, www.rotronic.fr
ROTRONIC Instruments (UK) Ltd, Crompton Fields, Phone +44 1293 571000, www.rotronic.co.uk

12.0816.0600

black

purple

yellow

orange

ROTRONIC Instrument Corp, NY 11788, USA, Phone +1 631 427-3898, www.rotronic-usa.com
ROTRONIC South East Asia Pte Ltd, Singapore 339156, Phone +65 6294 6065, www.rotronic.com.sg
ROTRONIC Shanghai Rep. Offi ce, Shanghai 200233, China, Phone +86 40 08162018, www.rotronic.cn