UEi Test Instruments INF195 User Manual

Inf195, Controls and indicators, Displays and indicators operating instructions

Advertising
background image

D:S=1

2:1

6.67@80”

3@36”

2@24”

INF1

95

Infrared

Thermomete

r

Controls and Indicators

1. Trigger: Initiates measurements.

2. IR Sensor

3. Laser Pointer Beam

4. LCD Display: Temperature

5. LCD Display: Secondary (MIN, MAX, AVG, DIF,

Alarm, Emissivity, Probe)

6. Battery Compartment

7. Thermocouple Socket

Displays and Indicators

Operating Instructions

Taking Measurements

To take a temperature measurement using your INF195, you

simply point the aperture at an object (with or without using

the laser sighting) and pull the trigger. The object’s

temperature will show up on the display and update at a rate

of approximately 2 times per second.

There will be a delay of approximately one-second between

the time you initially pull the trigger and the time the display

comes on. The 60-second auto-hold initiates at the moment

you release the trigger.

NOTE:

This thermometer will automatically shut off if left

idle for more than 60 seconds, unless in PRB mode. When in

PRB mode, instrument will shut off if left idle for more than

12 minutes.

Follow these general guidelines to ensure you get the most

accurate readings possible:

• Be sure the measured object fills the “spot” seen by

the aperture. The distance to spot ratio for the INF195

is 12:1.

This shows the one-foot spot fitting within the one-foot tar-

get area. At this distance, and anything closer, the

target’s temperature will be accurately measured.

NOTE:

If the two-foot diameter spot includes unwanted

objects in the background that are not part of the

one-foot square target. The temperature of the

background objects will be figured in with the target’s tem-

perature and throw off your measurement.

• When comparing temperatures of similar objects

that are far away, take your measurements at the

same distance and angle to the target each time.

• When looking for abnormally hot or cold targets it

may be acceptable to include background objects

so long as the temperatures in the background

and your methods are consistent.

• Consider the emissivity of the objects you

are measuring.

• Prepare a surface for measurement. Infrared

thermometers measure only the outer surface of

an object. If emissivity is affecting the

measurement, or you have difficulty putting the

object in the sensors line-of-site, you may need to

prepare a surface that’s easy for the infrared

thermometer to read. A piece of masking tape is a

good target and it will rapidly take on the

temperature of the object it is attached to.

• Be aware that you cannot measure the

temperature of air between the Infrared

thermometer and an object. Air vents (registers)

are quick to take on the temperature of outlet air.

However, you must aim directly at the vent if you

are measuring outlet air temperature.

• Keep your infrared thermometer away from

strong electrical fields. When working near a

strong electrical field, like that under the hood of

your car, watch for unusual readings or an “over

load” indication. Often, you can move the

thermometer just a few inches to escape the

influence of the interference.

• Keep your INF195 within its use and storage

temperature range. Excessive heat or cold will

adversely affect the accuracy of your readings.

When the trigger is pulled the target’s

temperature will be displayed in a near real-time

mode (less than 1/2 second between

measurements). The temperature will remain

on the display for seven seconds after the trigger

is released.

Emissivity

Not all surfaces emit infrared energy at the same

level. A shiny surface will emit much less infrared

energy at a given temperature when compared to

a flat black surface. The INF195 has three preset

emissivity levels used to compensate for variances

in your target.

High (0.95) will work for most common surfaces

and is set as default for many infrared

thermometers. Medium (070) is best for oxidized

copper or rusty iron. Use Low (0.30) for surfaces

that emit less energy such as aluminum or brass.

Mode and Functions

Press the “MODE” push-button will scroll through the

following options:

: Emissivity data. (The default emissivity

is 0.95)

: Press the “MODE” push-button, then press

“ “ or “ “ push-buttons to set the

emissivity, then press the “MODE” push-

button again to confirm it. The emissivity

can be changed between 0.30, 0.70 and

0.95.

: Press the “MODE” push-button for the

maximum (MAX), minimum (MIN), different

between MAX and MIN (DIF) and average

(AVG) modes. During the measurement,

the special modes reading will be displayed

at the bottom of the display near the

mode icon.

: Press the “ “ or “ “ push-buttons to

change the “High Alarm” (HAL) or

“Lo Alarm” (LAL), then press the trigger to

confirm it. For example: when the reading

81˚F < LAL 81.1˚F, the low “Lo Alarm” icon

will flash and you will hear a beep sound.

: Connect the thermocouple to the

thermocouple socket and apply the probes

to the target being measured. The

thermometer will display the temperature

automatically without pressing any buttons.

To see the minimum or maximum data

during the probe measurement, please hold

down the “ “ or “ “ push-buttons.

1-800-547-5740 • Fax: (503) 643-6322

www.ueitest.com • email: [email protected]

INSTRUCTION MANUAL

Introduction

The INF195 is perfect for frequent use in advanced or
specialty application environments, where wide tem-
perature ranges and superior optics are essential for
use on targets at greater distances. This full-featured
IR Thermometer is the most comprehensive and
best-valued Infrared Thermometer on the market.

The INF195’s increased features offer advanced trend
analysis for professionals that require a greater quality
of infrared optics, extended measurement and the ver-
satility of a K-Type surface or contact probe and three
level emissivity to simplify operation.

Features include:

• –76°~ 1022°F (-60˚~ 550˚C)
• 12:1 Distance to Spot Ratio
• MIN/MAX/Differential, Average
• 3 emissivity levels: (0.30, 0.70, 0.95)
• Hi/low Alarm (User adjustable)
• K-Type thermocouple input
• Carry Case

Safety Notes

Before using this meter, read all safety information
carefully. In this manual the word “WARNING” is used to

indicate conditions or actions that may pose physical haz-
ards to the user. The word “CAUTION” is used to indicate

conditions or actions that may damage this instrument.

NOTE:

The INF195 is not recommended for use on

shiny surfaces such as chrome, mirrors or

polished metals.

WARNING!

To avoid thermal shock, the instrument should be
stored at room temperature between 32˚ to 122˚F (0˚
to +50˚C).

WARNING!

DO NOT look directly into the laser beam. Permanent
eye damage may result.

1

4

5

2

3

6

7

INF195

1. Mode -button
2. Lock and Backlight

( ) Push-button

3. ˚F/˚C and Laser

( ) Push-button

4. Polarity

5. Main Display
6. Secondary Display
7 F˚ / C˚ Indicator
8. HOLD
9. Battery Life

D:S=12:1

6.67@80”

3@36”

2@24”

INF195

Infrared Thermomete

r

3

4

1

9

8

6

5

2

6

12