4BasicOperation

Connecting Cables

WARNING

Take the following precautions to avoid risk of fire or electric shock:

- Do not handle the power plug with wet hands.
- Use only the power cord provided with the instrument.

Connect the cables carefully so there are no problems operating the instrument.

CAUTION

Always turn off the power and unplug the power cord before connecting the cables of any

external devices to the TX-20.

Use the computer and other equipment that conform to the system standard IEC 60601-1 or

IEC 60950-1 for the Full Auto Tonometer TX-20. Be sure that the entire system conforms to
IEC 60601-1-1. Be sure to also use an isolation transformer conforming to IEC 60601-1 when a
computer conforming to IEC 60950-1 is used.
Otherwise, electric shock may occur. Ask your sales representative or local Canon dealer to
connect it to a computer.

If the LAN connector is being used to connect to a network, connect an isolation transformer for

networks between the TX-20 and network devices (HUB etc.) and between network devices.
If this is not done, there is a risk of short circuits or malfunctioning devices. Ask your sales
representative or local Canon dealer to connect it to a network.

1

Make sure that the power switch is off.

2

Connect a USB cable to input patient IDs with USB devices.

Remove the paper from the printer before turning the TX-20 on its side to attach cables. Otherwise
the paper may jam.

Always turn off the power and unplug the power cord before connecting the USB cable to the

TX-20.

For details on USB devices for inputting patient ID, contact your sales representative or local

Canon dealer.

USB cable