Common settings in measurement modes – Canon TX-20 Full Auto Tonometer User Manual

Page 61

Advertising
background image

5TypesofMeasurements

61

Common Settings in Measurement Modes

In the common settings menu, you can set auto-data output, alignment mode initial settings,
measurement results display order, measurement unit recovery position, distance moved when
switching eyes, and operation settings of the operation lever.

How to Enter the Common Settings Menu

MENU screen >

Measure tab > [Common]

Setting items

Description

Factory

settings

Auto Output Select whether to print (or output data) results automatically after

measuring both eyes, in full auto-alignment mode.

[ON]: Results are printed automatically.
[OFF]: Results are not printed automatically.
If the [Output] setting is [OFF], printing is not done even if this setting is [ON]
(see page 63).
If the [Interface] setting is [OFF], data is not output even if this setting is [ON]
(see page 64).

ON

Alignment
Mode

Select an alignment mode after initialization.

The main unit performs initialization in the following situations.
The power is turned on.

The PRINT button is pressed.

The CLS key is pressed.

The TX-20 recovers from power saving mode.

[FULL AUTO]: Full auto mode
[AUTO]: Auto mode

FULL
AUTO

Order

Select the order in which measurement results are shown.

[BY TIME]: Show in order of time.
[BY RELIABILITY]: Show in order of reliability.

BY TIME

Default Side Set the right eye side or left eye side as the recovery position for the

measurement unit.

The measurement unit moves to the recovery position of the side set in
the following situations.
Power on

After pressing the PRINT button

After pressing the CLS key

[RIGHT]: Right eye
[LEFT]: Left eye

RIGHT

R/L Distance Set the distance moved when switching between left and right eyes.

[58 to 66]: Distance moved (mm)

62

Joystick
Movement

Set whether to put the operation lever into proximity limiter mode.
In the proximity limiter mode, coarse adjustment motions are prevented
when the measurement unit moves within 2 mm of the alignment end
position. Beyond that, fine adjustment motion takes place until the unit
reaches the position.

[NORMAL]: Normal mode
[SAFETY]: Proximity limiter mode

NORMAL