5.4 OPTION Photography Mode

In this mode, OPTION photography using the exciter filter for fluorescein angiography is possible. The
control software must be set up in order to perform OPTION photography. For details on the settings, refer
to the operating manual for the control software.
The steps with no explanation in the procedure below are the same as in the color photography mode.
Refer to the procedure for the color photography mode (see page 28).

1

Enter the study information into the computer.

2

Select the OPTION photography mode.

Select OPTION with the MODE switch.
Or, select the photography mode in the control
software.
The COLOR and FLUO mode indicator lamps light up.

3

Disinfect the forehead rest and replace the chin rest paper.

4

Have the patient sit down.

5

Align the height of the patient’s eye.

6

Align the camera to the eye to be photographed.

7

Align the camera position.

8

Adjust brightness in field of view for observation.

9

Determine the area to be photographed.

10

Focus on the patient’s eye.

11

Select the photography range.

12

Take the image.

13

Review the image.

14

End the study.

15

Lock the stage unit.

16

Turn OFF the power of the camera main unit.