Myron L 6Pfc and 4P User Manual

Page 31

Advertising
background image

27

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the CELL ch location.

The display will show “Auto oFF” (see Figure 24, pg. 26).

4.

Press

to initiate. “CAL

Figure 25

CAL

will be displayed along with

15 SEC” or current Auto

oFF value (see Figure 25).

5.

Press

or

to

change the amount of time

Figure 26

CAL

(see Figure 26). Maximum

time is shown.

6.

Press

to accept the

change (new time).

XV.

USER MODE CALIBRATION LINC™ FUNCTION

The Linc™ function allows easy calibration when in User mode and the

user does not have a user standard solution to calibrate the instrument.

This function will ensure more repeatable and accurate measurements

than many other calibration methods. It is recommended that this

function be used to provide the highest degree of confidence when the

Ultrameter II is used in User mode. When Linc is used, the User mode

is linked to another standard, i.e., if User and KCl are linked, a KCI

standard solution is used to calibrate the instrument. It is that simple.

A. Calibration of Ultrameter II for use in User mode

1.

Press

or

key.

2.

Calibrate the unit using a Standard Solution (ref.

CaLIBRaTION, pg. 15).

3.

Place the Ultrameter II in User mode (ref. SOLUTION

SELECTION, pg. 12).

4.

Verify/Set the calibration linc. (See below – Setting User

Mode Calibration “Linc”.)