Myron L 6Pfc and 4P User Manual

Page 33

Advertising
background image

29

measurement keys will exit without changing the setting. User

mode “Linc” is now complete. The User mode will now use the

calibration gain constant used for the calibration of the

Standard Solution as outlined above.

C. Canceling User mode Calibration “Linc”

The Ultrameter II must be in “User” linked mode in order to cancel the

“Linc” (ref. SOLUTION SELECTION, pg. 12).

1.

Press “Linked” measurement key

,

or

.

Two solution icons will be shown in the left side of display -

User” and another, e.g., “KCl”.

2.

Press

key until the menu “Linc” appears (see Figure 27).

3.

Press

key; the instrument will display both “SEL” and

the “User” Icon.

4.

Press

key until “User” is the only solution icon being

displayed.

5.

Press

key.

6.

The User mode calibration “Linc” has now been canceled.

NOTES:

1.

To maintain repeatability, use the same standard solutions for

future calibrations.

2.

Calibration of the Ultrameter II Gain Factor for User mode is

not available when the calibration linc has been established.

The other calibration functions (i.e., Temperature Compensation

%/C settings and TDS Ratio settings) are still intact. To

perform a calibration of the User mode as described in User

Calibration Conductivity/TDS, pg. 16, the User mode Linc should

be canceled. See above, “Canceling User mode calibration

“Linc””.

3.

Once a “Linc” has been established for User mode , the “Linc”

will apply to all measurement modes using User solution

selection (i.e., TDS/User, Cond/User or Res/User).