Myron L 6Pfc and 4P User Manual

Page 42

Advertising
background image

38

Chart 1

0 5 10 15 20 25 30 35 40 45 50 55 60

1.500%

1.600%

1.700%

1.800%

1.900%

2.000%

2.100%

2.200%

2.300%

2.400%

2.500%

KCl % / °C

% / °C

Temperature

C. an Example of 2 different solution selections and the

resulting compensation

How much error results from treating natural water as if it were KCl at

15°C?
a tap water solution should be compensated as 442 with a tempco of

1.68 %/°C, where the KCl value used would be 1.90 %/°C.
Suppose a measurement at 15°C/59°F is 900 microsiemens of true

uncompensated conductivity.
Using a 442 correction of 10 (degrees below 25) x 1.68% indicates the

solution is reading 16.8% low. For correction, dividing by (.832) yields

1082 microsiemens as a compensated reading.
a KCl correction of 10 (degrees below 25) x 1.9% indicates the solution

is reading 19% low. Dividing by (.81) yields 1111 microsiemens for a

compensated reading. The difference is 29 out of 1082 = 2.7%.

D. A Chart of Comparative Error

In the range of 1000 µS, the error using KCl on a solution that should

be compensated as NaCl or as 442, is illustrated in Chart 2 on pg. 39.

Users wanting to measure natural water based solutions to 1% would have

to alter the internal compensation to the more suitable preloaded “442”

values or stay close to 25°C. Users who have standardized to KCl- based

compensation may want to stick with it, regardless of increasing error as

you get further from 25°C. The Ultrameter II will provide the repeatability

and convertibility of data necessary for relative values for process control.

Advertising