5.2. Levelling the scale

Level scale by means of turning levelling feet and external level indicator.
Put the pads under the regulation legs and observe the level condition
indicator. The instrument is levelled correctly if the air bubble do not
move more than 2mm from the central position.

Each leg can be screwed in and out in order to level the scale. Correct
levelling can be done by putting steel pads under the legs.



6. GETTING STARTED

After unpacking and mounting (chapter 5 of this manual):

• Turn the device on using the

key – keep pressing the key

for about 0.5 sec,

• Wait for the test completion,
• Then you will see zero indication and pictograms:

- zero indication

- stable result

kg

- weight unit

• If the indication is not zero press zero key.