19.6. Continuous transmission

The indicator can work in a continuous transmission mode. It can
be switched on or off in parameters or using RS232 commands.

The frame format sent by the indicator in case of setting
<P2.Prnt> to CntA:

1 2

3

4

5

6 7-15

16 17 18 19 20 21

S I space stability space sign mass space

Unit

CR LF

Stability character

[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range

sign

[space] for positive values or
[-] for negative values

mass

9 characters justified to the right

unit

3 characters justified to the left

command

3 characters justified to the left



The frame format sent by the indicator in case of setting <P2.Prnt>
to Cntb:

1 2 3

4

5

6 7-15 16 17 18 19 20 21

S U I

stability

space sign mass space

unit

CR

LF



19.7. Configuring printouts

General information

If some information included are redundant or not sufficient and there
is a necessity of changes one can design their own protocol format in
EDYTOR WAG computer program. This piece of software is accessible
in: http://www.radwag.com