Zeroing, Weighing for two range scales – RADWAG PUE C indicator User Manual

Page 21

Advertising
background image

EXI-01-04-06-14-GB

- 21 -

NOTICE:
Tarring by inscribing tare value cannot be performed when the tare value
has already been implemented into the scale memory. In such a case the
display shows <Err3> message, short sound signal is heard.


12.3. Zeroing

In order to zero the indication press zeroing key -

.

Zero value is shown

on a display next to symbols:

and

.


The new zero point is treated by the scale as the precise zero. Zeroing is
possible while stable status is shown on the display.


NOTICE:
Display status zeroing is possible only in range up to ±2% of maximum
scale capacity. If zeroed value is bigger than ±2% of maximum capacity,
display shows Err 2 message, short sound signal is heard.


12.4. Weighing for two range scales

Switching from the I range to the II range is carried out automatically after
exceeding the switching point value (Max of the I range).

Weighing in the II range is signalled by a pictogram, located in the upper
left-hand corner of a display.

After unloading, the scale returns to zero. Weighing in the II range takes
place until unloading the pan.


II range weighing automatically turns to I range weighing after taking the
load off the weighing pan when the scale returns to AUTOZERO zone – the

pictogram is seen.