13.2. Autozero


„AUTOZERO” function has been introduced in order to ensure precise
indication. The intended use of this function is automatic control and
correction of zero indication.

When function is active comparison of results takes place at specified time
intervals. If results differ by value smaller than declared AUTOZERO range,
e.g. one division, scale is zeroed automatically, stable state is marked –

and zero indication –

is displayed.


When AUTOZERO function is enabled, each result starts from precise
zero. However there are some cases when this function can disturb
measurement, e.g. very slow process of load depositing on the pan (load
pouring). In such case, correction system for zero indication can also
correct indication of real mass of load.

Procedure:

• Following point 11 of this manual enter <P1.rEAd> submenu, next:

AUTO

no

- autozero disabled

AUTO

YES - autozero enabled

Return to weighing:

See point - 11.2. – return to weighing.