57

7.

Reports

Examples of each report available in the SAM 935 are included here with an explanation of their contents.

7.1

Area Monitor Report of Alarms

The area monitor report lists information about alarms. You get this report by pressing F3 while in review
mode on the monitoring display (if F3 is labelled P

RN

S

EL

). Only the stored alarm that is currently displayed

will be printed.

7.2

Background Report

The background report is a simple report since it gives only system setup information and gross counts. It
is useful to get a “full” report, which prints out the background spectrum. Normally there should be no
peaks in a background spectrum. The presence of peaks indicates either contamination in or on the detec-
tor or radioactive material in the vicinity of the detector whose radiation are not being stopped by whatever
shielding may be present.

Area Monitor Report

Alarm # 1 at 14-Jul-1999 21:14:44 for 21 sec

Nuclide(s) Peak Conf Total Conf Peak Dose Avg Dose

Cs137 20978.87 28429.04 3.11 mR/hr 2.42 mR/hr

Full 15176.49 20573.70 3.11 mR/hr 2.42 mR/hr

I125 11210.00 15084.15 0.00 nR/hr 0.00 nR/hr

The following monitors had no measured activity:

Am241 Ba133 Co57 Co60 Ga67 I131 In111 K40 Pd103 Ra226 Tc99m Tl201 U235

Xe133

Neutron counts/min: 292

East Tennessee Medical Center

BACKGROUND REPORT

DATE: 31-Aug-1995 21:29 BIAS: 795

EN CAL DATE: 31-Aug-1995 21:29 COARSE GAIN: 1

No Efficiency Calibration FINE GAIN: 1.31

BKG DATE: 31-Aug-1995 21:33 LOW DISC: 0.82%

GROSS CPM: 698 HIGH DISC: 105.00%

ELAPSED LT: 180.00

GROSS INTEGRAL: 698 ELAPSED RT: 180.02

DEAD TIME: 0.01%

FULL SCALE: 23