Eppendorf New Brunswick Shaker Operating Key User Manual

New brunswick™ shaker operating key

Advertising
background image

www.eppendorf.com

To set and start a timed run:

1. Press SELECT until HRS lights.
2. Press

or to set time (between 0.1 HRS and 99.9 HRS).

3. While SET is lit, press START/STOP. TIME will light and Display will

show time remaining.

To cancel:

1. Press SELECT to HRS.
2. Press

or , then START/STOP. Time light turns off. Display reads

“Off”.

Audible alarm:

1. Press SELECT to acknowledge. Stops sounding, but
remains enabled.
2. To deactivate, SELECT to HRS, then press

and

SIMULTANEOUSLY , then press START/STOP. MUTE will light.
3. To reactivate, SELECT to HRS, then press

and

SIMULTANEOUSLY, then press START/STOP. MUTE will go off.

Total running time:

1. Press SELECT to HRS.
2. Press

and SIMULTANEOUSLY.

3. Press START/STOP. Display indicates hundreds of hours.

Flashing display indicates power interruption.
Press any key to deactivate.

NEW BRUNSWICK™ SHAKER OPERATING KEY

M1190-9600 Revision C

12. MAINT: Every 10,000 hours of opera-
tion, lights to signal need for preventative
maintenance

11. SET: For establishment of setpoints,
lights when using or key

10. TIME: Lights during a timed run.
Flashes at end of run.

9. MUTE: Lights when audible alarm is
deactivated.

8. Starts or stops shaking.

7. Increases setpoint for agitation speed
(RPM), time (HRS) or temperature (°C).

1. RPM: Lights when display shows
speed, i.e., actual speed or setpoint in
SET mode. Flashes when 5 RPM above
or below setpoint.
2. HRS: Lights when display shows
time, i.e., actual time remaining, or
setpoint in SET mode, or total running
time.
3. °C: Lights when display shows
temperature, i.e., actual chamber or
bath temperature, or setpoint in SET
mode. Flashes when 1 °C above or
below setpoint.
4. *: Innova 3100 - Flashes when water
level is below setpoint.

5. Determines the Function display:
RPM, HRS, °C

6. Decreases setpoint for agitation speed
(RPM), time (HRS) or temperature (°C).

To adjust running time:
While TIME is lit, press or to set a new time.

1
2

3
4

5

6

7

8

12
11

10

9

SELECT

START

STOP

MAINT

SET

TIME

MUTE

RPM

HRS

° C

*

Advertising