DESCRIPTION

MaxiThermal software gives users of Logger-Trac family

devices a simple yet flexible interface for setup and data

retrieval. Downloaded data from Logger-Trac™ devices

may be displayed graphically, in table format, or downloaded

in CSV format for spreadsheet viewing and analysis. The

software is compatible with Windows XP, Vista, 7 and 8.
SOFTWARE INSTALLATION

• Run provided setup package and follow the screen

prompt to complete installation

• Restart computer
HARDWARE SETUP

• For Logger-Trac Temperature:

o Insert device into USB cradle

o Connect cradle to computer using USB cable

• For Logger-Trac Humidity or Logger-Trac LN2 connect

device directly to computer using USB cable

• If it is the first time connecting to computer, USB driver

will install automatically

• Run MaxiThermal Software by clicking icon on desktop
DATA-LOGGER SETUP

Logger-Trac™ devices require initialization before each

use. Initialization sets up all necessary parameters that the

device requires.

• Click Logger Initialization tool bar, or click Logger ->

Logger Initialization. Logger initialization window opens.

• (Opt.) Enter Description that describes the shipment

• Window displays current computer time and time zone

• (Opt.) Select time zone for a drop down list. Default is

the same as computer setting

• (Opt.) Check/uncheck Auto Adjust Daylight Time

• Select temperature measurement unit

• (Opt.) Specify start delay in minutes from drop-down list

• Specify recording duration up to maximum 365 days.

The software automatically adjusts sample interval based

upon the entered Duration time

• Specify sample interval. Software automatically adjusts

recording duration based upon entered sample interval

• Specify Memory Configuration

o Record to End of Memory – Data recording halts once

memory is full

o Continuous Memory – Data recorded overwrites pre-

viously recorded data once memory is full beginning

with oldest data

• Specify Alarm settings

o Maximum Alarm limit

o Minimum Alarm limit

o Quarantine value (Logger-Trac Temperature only) –

To enable the Quarantine feature, set to a nonzero

integer. Quarantine count increments when either

the temperature transitions to outside the alarm limits

or when a temperature reading is logged outside

the defined alarm limits. Once the count increments

greater than or equal to the set quarantine value, the

QUARANTINE LCD symbol becomes active. This

feature gives the user an approximation for the cumu-

lative time temperature was outside the alarm limits.

Interrogate logged data for further analysis of results.

o No alarm checkbox

• Enable/Disable Stop Key. If Yes is selected, logger can

be stopped by pressing and holding Start button by

user after it starts logging data. If No is checked, logger

cannot be stopped once it starts using the Start button

• Once all settings are configured as desired, click the

Erase all data button to erase old settings and data from

the device.

• Click Write to device button to program device with new

settings

• Close logger initialization window
DATA RETRIEVING

Once shipment is complete, recorded data can be download-

ed to computer for viewing and analysis.

• Connect Logger-Trac to computer, reference Hardware

Setup section above.

• If Logger-Trac is currently logging, click Logger -> Stop

C\TempLCD(TYPE A) to stop the data logging

• Click Read Logger tool bar, or menu Logger -> Read

Logger

• When prompted, specify a location and file name to save

recorded data.

• Once data saving is complete, the software window

shows the graph display of the recorded data

EXPORT DATA

Export data is to export saved data that has been download-

ed from the Data-Logger to CSV file in order to view data in

Microsoft Excel or similar

• Click Export tool bar, or click File -> Export data

• When prompted, enter file name with extension .csv
DELETE DATA

• While connected, from the Logger Initialization screen,

click Erase all data. All data recorded in the Data-Logger

will be erased. However, data that has been read and

saved to the computer will not be deleted.

VIEW DATA

• Once Read Data from the logger is complete, and graph

is displayed on screen, click View data tool bar, or click

Graph -> View data to view recorded data with time

stamp

• To view previously saved data, click Open tool bar, or

click File -> Open to open the Open window

• Select previously saved data file to open
DATA

Recorded data cannot be deleted without using MaxiThermal

software.
DETAILED ELECTRONIC DATA-LOGGER MAXITHERMAL

SOFTWARE INSTRUCTIONS

After Maxithermal Software is opened, the screen is shown

on Fig. 1, and toolbar function is labeled in Fig. 1.

FIGURE 1

1. File

1.1. Open

• View previously saved data file. Data files have

extension .Mp_Lgr2

1.2. Export data

• Export data in CSV format so that data can be

viewed in Microsoft Excel or similar

1.3. Preferences

• Set software preferences: Graph (Fig. 2), Language

(Fig. 3), Defaults (Fig. 4),, Directory (Fig. 5)

FIGURE 2

FIGURE 3

FIGURE 4

FIGURE 5

2. Logger

2.1. Logger initialization – main screen to set up Data-Log-

ger (Fig. 6)

FIGURE 6

2.1.1. Model Number

• Displays the connected Data-Logger model number

2.1.2. Description

• Enters a meaningful description for the shipment.

Maximum number of characters is 20, and cannot have

<>?/:|\*

2.1.3. Current Clock

• Displays computer’s current time and time zone

2.1.4. Logger Clock

• From the drop-down list, select the time zone for Da-

ta-Logger. Default is the same time zone as computer’s

2.1.5. Auto Adjust Daylight Time

• Check the box to enable auto-adjust daylight saving

time; uncheck the box to disable the function

• The displayed date and time will be auto-adjusted

according the setting above

2.1.6. Unit

• Check the appropriate box to select temperature unit, in

Celsius or Fahrenheit

2.1.7. Start Delay

• Specify the delay time from the drop-down list. The

Data-Logger will not begin to record data until the delay

time has been reached after Data-Logger starts

2.1.8. Duration

• Specify the duration during which Data-Logger records

data. Maximum duration is 365 days. Once duration is

set, the software will adjust Sample Interval automati-

cally

• User can also select duration from drop-down list

2.1.9. Sample Interval

• Specify the interval that the Data-Logger samples. The

longer the sample interval, the longer the duration.

2.1.10. Memory Configuration

• Record to End of memory

• Continuous Memory

2.1.11. Alarm

• Maximum: set the upper limit

• Minimum: set the lower limit

• Quarantine: (For Logger-Trac Temperature only) The

number of recorded temperature readings outside the

alarm limits to trigger Quarantine LCD symbol.

• No alarm: If checked, Alarm function will be disabled,

and Maximum, Minimum, Quarantine will be grayed out

2.1.12. Stop Key

• Yes: If checked, user is able to stop the Data-Logger

from logging data by pressing the Start key.

• No: If checked, Data-Logger cannot be stopped from

logging once it begins to log. It can only be stopped

with MaxiThermal Software/USB connection.

2.1.13. Write to device

• Upload all user settings to Electronic Data-Logger

• If Data-Logger is to be reinitialized for next shipment,

click Erase all data first. The software will prompt a

warning message. See Fig. 7.

FIGURE 7

Advertising