Ie i, A. set-up – Heidolph Pump Heads User Manual

Page 46

Advertising
background image

I

E

I

46

Turn power switch to OFF when item is out of use or before disconnecting from

mains.

Repair work is limited to skilled personnel authorized by Heidolph.

Your item needs a solid stand.

Before turning pump drives ON, make sure your hands are clear of the squeeze

rollers; contusion hazard !

A. SET-UP


1. General (fig. 21)

- When using more pump drives at the same time (2 ea. max.),

they may be stacked to save space.


2. Install peristaltic pump heads on pump drives

2.1. SP quick (figs. 22, 23)

- Press SP quick pump head (3) against

PD.... pump drive front panel so that thread

(4) and coupling (5) are in line with

threaded bolt and coupling on pump head.

- Protruding parts of the coupling crown need

to slip into cut-outs provided in pump drive

coupling stud. To this end tilt pump head

around rotor center or move rotor with

forefinger, hose bed being open (refer to

chapter 3.3).

-

When coupling is seated properly, secure

pump head (3) with 2 ea. knurled-head

bolts (6) by hand; if required, tighten with

screwdriver or coin.

-

fig. 21

3

6

5

4

fig. 22

6

fig. 23

Advertising
This manual is related to the following products: