I

E

I

49

3.4. SP standard (fig. 26)

-

Loosen knurled nuts (22 & 23) to remove

clamping lever (20) and cover (21). Seat

tube (17) in tube bed (180 degrees loop),

rotate rotor (24) by hand until hose (17) is

pulled into the gap between the rollers (25)

and tube bed (26).

-

Check tube being squeezed poperly by

rotating the rotor (24) through several

revolutions.

-

Install clamp lever (20) and cover (21); hold

both tube ends tight while tightening knurled

nut (23).

-

Holding tube ends tight over the rollers (25)

will make sure both rollers (25) are centered

over the tube (17) which is an essential

requirement for proper feeding action and

standstill sealing.

-

Do a short test run (function test).

3.5. SP vario (fig. 27)

-

The SP vario pump head accepts tubes of different wall thicknesses. Adjustment uses

knurled knob (27).

-

Set clockwise rotation with selector (31), refer to figs. 47, 49 and 51

-

Uninstall clamp lever (20), cover (21) by loosening knurled nuts (22 & 23).

-

Set clearance to max. tube thickness with knurled knob (27); rotate in ”+“ direction

-

Introduce a 180 degrees tube (17) loop in tube bed, rotate rotor (24) by hand until tube (17)

is pulled into the gap between rollers (25) and tube bed (26).

-

Run rotor (24) through several revolutions until tube is squeezed over its entire width.

-

Energize pump on condition.

fig. 26

26

21

24

22

17

23

20

25